PUMP TUBE REPLACEMENT
19
11. PUMP TUBE REPLACEMENT
To remove the tube of the peristaltic pump follow next steps:
y
Detach the old tube system from the reagent bottle.
y
Grasp one fixing ring of the peristaltic pump tube.
y
Pull the tube until it’s taken out from its location.
y
Remove the other side of the tube.
To mount the new peristaltic pump tube follow next steps:
y
Position one peristaltic pump fixing ring on its location.
y
Stretch the tube over the peristaltic pump cylinders.
y
Fix the second pump fixing ring on its location.
y
Attach the tube to the reagent bottle.
Note:
Purge the peristaltic pump until drops of reagent appears on the dosing tip by
pressing the PUMP button.