Always  remove  the  probe’s  protective  cap  before  taking  any 

measurement.  Make  sure  that  the  meter  has  been  calibrated 

before taking any measurements

Immerse the tip (2.5 cm) of the EC probe into the sample. (for 

example it can float with the help of a piece of polystyrol).

Turn the instrument on by pressing the ON/OFF button.

The nature of the setpoint can be selected by setting the switch 

to the desired position (ABOVE or BELOW). 

USE ALWAYS BELOW!

Allow the reading to stabilize and the meter will start continuous 

monitoring.

A  blinking  alarm  will  indicate  when  the  measured  EC  value 

is higher or lower than selected setpoint, depending on user  

selection.  When  the  alarm  is  activated,  it  also  activates  the 

 

connection base and the auxiliary equipment.

The setpoint can be selected by adjusting the central front knob 

to the desired value. The selectable range is from 0.8 to 2.8 mS. 

Connect the supplied 240VAC/12VDC power 

adapter to the meter and to the main line.

Connect the connection base (black and white 

outlet) to the main line.

INSTALLATION OF CONTROLLER:

CALIBRATION:

Clean the probe with alcohol and let it dry (if a more thorough

cleaning is required, brush the metal pins with fine sandpa

-

per and then rinse with water).

Open a sachet of suitable conductivity calibration solution and

immerse the probe, making sure that metal pins are completely

submerged. Use M10031 (1413 μS/cm).

Summary of Contents for MC745

Page 1: ...30VAC 1A 50Hz pH electrode MA812 2 included Environment 0 to 50 C 95 RH not condensing Power supply 12 VDC included Dimensions 148 5 x 82 5 x 32 mm Weight 180 g MP810 Peristaltic Dosing Pump Constant Dosing flow of 1 5 l h 25 cc min No regulation Maximum pressure 1 bar Made of Polipropilene PP Peristaltic tube in Santoprene Power supply 230 VAC 10 Authorized Distributor in Australia Pacific Sensor...

Page 2: ... higher or lower than selected setpoint depending on user selection When the alarm is activated it also activates the connection base and the auxiliary equipment The setpoint can be selected by adjusting the central front knob to the desired value The selectable range is from 0 8 to 2 8 mS Connect the supplied 240VAC 12VDC power adapter to the meter and to the main line Connect the connection base...

Page 3: ...y to inject into a piping This fitting has a thread that links it to the piping The calibration is now complete and the meter is ready for use The instrument should be re calibrated at least once a month or when the probe is changed Mount the bracket in the wall in upright position above the water container which contains a solution of EC reducing acid and water Place the pump in vertical position...

Page 4: ...roneous operation by the user please return the parts to either dealer or our office and the repair will be effected free of charge Note We reserve the right to modify the design construction and appearance of our products without ad vance notice IMPORTANT For a smooth adjustment of the EC of your nutrient mix it is advisable to dilute each liquid nutrient with water in a jug Example For a 100 lit...

Reviews: