OPERATION

 

y

Connect the supplied 12VDC power adapter to the 
meter and to the mains line.

 

y

Connect the pH-electrode to the BNC socket on the 
bottom of the meter.

 

y

Always remove the electrode protective cap before 
taking any measurement. If the electrode has been 
left dry, soak the tip (bottom 2.5 cm) in 

M10000B

 

rinse solution for a few minutes to reactivate it.

 

y

Make sure that the meter has been calibrated 
before taking any measurements (see Calibration 
Procedure).

 

y

Immerse the tip (2.5 cm) of the pH-electrode into 
the sample.

 

y

Turn the instrument on by pressing the ON/OFF key.

 

y

Allow the reading to stabilize and the meter will 
start continuous monitoring.

 

y

A blinking alarm will indicate when the measured 
pH value is higher or lower than selected setpoint, 
depending on user selection.

Note:

 The output power contact (MC122 only) has no 

protection fuse inside the meter. It is recommended to 
protected it outside, against failure.

SETPOINT

 

y

The setpoint can be selected by adjusting the 
central front knob to the desired value.

 

y

The selectable range is from 3.5 to 7.5 pH for 

MC110

from 5.5 to 9.5 pH for 

MC120

 and 

MC122

.

 

y

The nature of the setpoint can be selected by 
setting the switch to the desired position (ABOVE 
or BELOW).

CALIBRATION PROCEDURE

 

y

Remove the protective cap from the electrode.

 

y

Immerse the electrode tip into a new sachet of 
pH 7 calibration solution and allow the reading to 
stabilize.

 

y

Adjust the pH7 calibration trimmer (on the front) to 
display “7.0 pH”.

 

y

Open a new sachet of pH 4 calibration solution and 
use a small quantity to rinse the electrode.

 

y

Immerse the electrode tip into the pH 4 sachet and 
allow the reading to stabilize.

 

y

Adjust the pH4/10 calibration trimmer (on the front) 
to display “4.0 pH”.

 

y

The calibration is now complete and the meter is 
ready to take measurements.

 

y

It is suggested to recalibrate the meter at least once 
a month, after a prolonged stocking time and after 
pH-electrode replacement.