Millipore Amicon 6028 User Manual Download Page 4

Setup

Reservoir Assembly/Disassembly

1.  Remove pressure relief valve (if installed) and slide reservoir out of retaining stand. 

Disassemble and clean all components prior to first time use.

2.  Attach a 5 in. (12.7 cm) piece of ¼ in. (6.4 mm) outer diameter (OD) tubing to the liquid 

outlet port (labeled "L") on the bottom of the reservoir cap.

Liquid port 

tubing

Liquid outlet port (L)

Gas port (G)

Pressure relief valve

Sample port

Cap

Retaining stand

Base

Reservoir O-ring

Reservoir O-ring

Pressure relief  

valve O-ring

Reservoir body

3.  Ensure that the reservoir O-rings are positioned correctly in the grooves of base and cap. If 

needed, lubricate O-rings lightly with water or petroleum jelly. Do not allow petroleum jelly 

to come in contact with sample. 

NOTE: 

Inspect O-rings for cuts or wear; if necessary, replace with spare O-rings.

4.  Push base and cap into reservoir body until properly seated, then slide reservoir assembly 

into retaining stand.

WARNING: 

Never use reservoir without its retaining stand! The cap will pop off when 

reservoir is pressurized and liquid in reservoir may splatter.

Inlet/Outlet Tube Fitting Assembly

The polyethylene tubing required for assembling the reservoir with the stirred cell and selector 

valve is supplied with the following products. Tubing designations listed below refer to the 

ports on the Amicon® Stirred Cell Selector Valve.
RES IN and RES OUT port tubing is supplied with the Amicon® Stirred Cell Reservoir. 
CELL tubing is supplied with the Amicon® Stirred Cell (quick-connect fitting on one end).
GAS SUPP tubing is supplied with the Amicon® Stirred Cell Selector Valve.

NOTE: 

Attachment to older model stirred cells is the same, except that the cell cap connection 

is the tubing assembly shown on the next page, rather than quick-connect.

Operation

WARNING: 

Do not exceed pressure limit of 5.2 bar (75 psi). During diafiltration, the pressure 

should not exceed 3.8 bar (55 psi), in order to maintain equilibrium between the 

stirred cell and the reservoir. If the liquid level increases slightly, concentrate 

briefly.
To prevent fluid from entering the stirred cell pressure relief valve during 

diafiltration, do not exceed the maximum working volume  

(e.g., 50 mL for the 50 mL stirred cell).

NOTE: 

Nitrogen gas is recommended for pressurizing the Amicon® Stirred Cell and reservoir.

Refer to Amicon® Stirred Cell Selector Valve user guide for detailed instructions on 

concentrating and diafiltering with the selector valve and reservoir.
1.  To concentrate, push the 

GAS spool valve plug on the selector valve in until spool plug seats 

fully. 

2.  Initiate stirring and slowly pressurize the reservoir. Check connections for leaks, then 

increase pressure as desired. Both the reservoir and stirred cell are pressurized, but the cell 

is "seeing" only gas and there is no flow of liquid from the reservoir to the cell.

3.  To diafilter, first run in 

GAS mode (steps 1 and 2) for 5–10 seconds, then push the LIQ spool 

valve plug in. This allows pressurized liquid to flow out of the reservoir and into the stirred 

cell.

4.  Proceed with concentration or diafiltration until desired sample volume or buffer exchange 

has been achieved.

Shutdown and Disassembly

1.  Turn off pressure at source, then turn off magnetic stirrer.
2.  Vent pressure from system using the reservoir pressure relief valve, rather than the stirred 

cell pressure relief valve. This prevents liquid remaining in tubing/valve between cell and 

reservoir from flowing into the cell.

3.  To disassemble, disconnect gas and liquid tube fittings.
4.  Remove pressure relief valve and slide reservoir out of retaining stand.
5.  Pry top and bottom caps from reservoir sleeve using cap removal tool provided.
6.  Remove O-rings from top and bottom caps if needed.

Cleaning and Maintenance

1.  After each use, wash the reservoir and all fluid-carrying tubing with mild laboratory 

detergent and rinse with deionized water.

2.  If tube fitting assemblies are removed for any reason, rewrap the threads with 

polytetrafluoroethylene tape before replacing them back into the cover. Use moderate force 

when replacing fittings.

Sterilization/Disinfecting

The reservoir is 

NOT autoclavable. To sterilize, use standard sterilizing gas mixtures. To 

disinfect, use 5% formalin. Do 

NOT use alcohol.

Additional tubing may be needed if using the reservoir with other devices.
1.  Lay out the stirred cell, selector valve, reservoir, and four pieces of tubing in positions 

appropriate for their use together. Estimate length of tubing required for each connection 

and cut to length if necessary.

2.  Carefully unscrew each of the four hex caps from the selector valve body and remove the 

O-ring. A grab ring and spacer are included inside each hex cap.

3.  Slide the hex cap and grab ring (flat portion facing 

hex cap) onto one end of each of the four pieces of 

tubing leaving no more than ½ in. (1.3 cm) of tubing 

beyond grab ring. Add a spacer (countersink toward 

grab ring) and O-ring onto the end of each piece of 

tubing.

4.  Attach each tubing assembly to the appropriate 

selector valve port and hand tighten the hex caps.

5.  Add appropriate tube fittings to the free ends of 

RES INRES OUT, and GAS SUPP tubing 

pieces. The 

CELL tubing already has a quick-connect fitting on it.

6.  Refer to the diagram below and connect the ends of the tubing as follows:

 

● Connect 

GAS SUPP to external pressure source.

 

● Connect 

CELL to quick-connect inlet fitting on stirred cell.

 

● Connect 

RES IN to inlet fitting (labeled "G" on Amicon® Stirred Cell Reservoir cap).

 

● Connect 

RES OUT to outlet fitting (labeled "L" on Amicon® Stirred Cell Reservoir cap).

7.  Assemble the Amicon® Stirred Cell according to the user guide and fill with sample.
8.  Add sample or diafiltration buffer to the reservoir through the recessed sample port. Install 

pressure relief valve, taking care not to cross-thread the valve in the sample port. 

200
175
150
125
100
75

60

50

40

30

20 ml

250

50

100

150

200

GAS

GAS

SUPP

CELL

RES IN

RES OUT

LIQ

100

800

700

600

500

400

300

200

Amicon® Stirred Cell 

Reservoir 

(cat. no. 6028)

Amicon® Stirred Cell 

Selector Valve 

(cat. no. 6003)

Amicon® Stirred Cell

To regulated  

pressure source

½ in.

Grab ring

O-ring

Spacer

Tubing

Hex cap

Setup,

 

continued

99137_Rev1015.indd   2

10/26/2015   4:06:15 PM