Milkotester BS - W8 Operation Manual Download Page 9

Milkotester Ltd 

Thermostatic device 

 

 

Model   

 

BS – W8 

 

Application 

Incubating antibiotic test for raw milk   

Compatible test/ tube 

Beta Star  , Beta Star combo 

Time range 

0 to 600 min. 

Temperature range 

30° C to 80°C 

Working modes 

Mode 1 

 

Temperature :  47° C  

 

Timer 1: 2min. 

 

Timer 2: 3min. 
 
 

Mode 2 

 

Temperature :  47° C  

 

Timer 1: 3min. 

 

Timer 2: 2min. 
 

 

Number of wells 

8  

Electrical parameters 

Switching adapter: 
Input 100-240 V~1.6A max , 50~60 Hz 
12V ~24V 

Dimensions ( W x L x H)  160 x 90 x 80 mm 

Weight 

1.005 kg 

 

Summary of Contents for BS - W8

Page 1: ...Dry incubator Operation manual www milkotester com www milkanalyzers eu ...

Page 2: ...ocal AC power supply The incubator placed on a firm surface Disconnect the thermostatic device from the power supply when cleaning it Service should be performed by authorized personnel only 1 Overview functions 1 Green led shows the status of the timer 1 also in programming mode 2 Green led shows the status of the timer 2 also in programming mode 3 Red led shows the status of the heater on off 4 ...

Page 3: ...n thru one minute The thermostatic device is factory set for temperature and time according to the different models ET W24 Temperature Timer 1 Timer 2 MODE 1 40 C 3min 3min MODE 2 65 C 180min DC W14 Temperature Timer 1 Timer 2 MODE 1 56 C 10min MODE 2 64 C 180min CH W12 Temperature Timer 1 Timer 2 MODE 1 56 C 10min 10min MODE 2 56 C 10min 10min BS W8 Temperature Timer 1 Timer 2 MODE 1 47 7 C 2min ...

Page 4: ...by keeping the button pressed 3 3 After 3 sec you will hear sound signal and number 1 will appear on the display 4 Setting of working mode 2 Enter working mode 2 4 1 The thermostatic device must be turned off from the power supply 4 2 Press the button DOWN and turn on the power by keeping the button pressed 4 3 After 3 sec you will hear sound signal and number 2 will appear on the display 1 2 3 1 ...

Page 5: ...By pressing the buttons UP and DOWN you can adjust the desired temperature Press again the button SET to save the settings and accompanied with double sound signal time correction will be done automatically At customer s request two working cycles can be set The indicator of Timer 1 will illuminate and the factory set time for cycle 1 will appear on the display By buttons UP and DOWN the time can ...

Page 6: ...ART as the indicator displays the specified time and start Timer 1 timer 1 indicator lights Time decrease and reaching 000 the device gives an intermitted sound signal and on the display is flashing 1 After pressing the START button starts Timer 2 which indicates time After time indication appears flashing symbol 2 and intermitted sound signal The device waits for pressing the START button to comp...

Page 7: ...ensor Time range 0 to 600 min Temperature range 30 C to 80 C Working modes Mode 1 Temperature 40 C Timer 1 3min Timer 2 3min Mode 2 Temperature 65 C Timer 1 180min Number of wells 24 Electrical parameters Switching adapter Input 100 240 V 1 6A max 50 60 Hz 12V 24V Dimensions W x L x H 160 x 90 x 50 mm Weight 0 620 kg Cover Box material Stainless steel ...

Page 8: ...e range 0 to 600 min Temperature range 30 C to 80 C Working modes Mode 1 Temperature 40 C Timer 1 1min Mode 2 Temperature 56 C Timer 1 3min Timer 2 8min Number of wells 2 Electrical parameters Switching adapter Input 100 240 V 1 6A max 50 60 Hz 12V 24V Dimensions W x L x H 190 x 90 x 60 mm Weight 0 880 kg Cover Box material Stainless steel ...

Page 9: ...ta Star combo Time range 0 to 600 min Temperature range 30 C to 80 C Working modes Mode 1 Temperature 47 C Timer 1 2min Timer 2 3min Mode 2 Temperature 47 C Timer 1 3min Timer 2 2min Number of wells 8 Electrical parameters Switching adapter Input 100 240 V 1 6A max 50 60 Hz 12V 24V Dimensions W x L x H 160 x 90 x 80 mm Weight 1 005 kg ...

Page 10: ...tube Delvotest Copan Time range 0 to 600 min Temperature range 30 C to 80 C Working modes Mode 1 Temperature 56 C Timer 1 10min Mode 2 Temperature 64 C Timer 1 180min Number of wells 14 Electrical parameters Switching adapter Input 100 240 V 1 6A max 50 60 Hz 12V 24V Dimensions W x L x H 160 x 90 x 80 mm Weight 1 145 kg Cover Box material Stainless steel ...

Page 11: ...ime range 0 to 600 min Temperature range 30 C to 80 C Working modes Mode 1 Temperature 56 C Timer 1 10min Mode 2 Temperature 56 C Timer 1 10min Number of wells 12 Electrical parameters Switching adapter Input 100 240 V 1 6A max 50 60 Hz 12V 24V Dimensions W x L x H 160 x 105 x 80 mm Weight 1 535 kg Cover Box material Stainless steel ...

Page 12: ...mage any part of the thermostatic device please follow the cleaning instructions below Clean the metal cover box of the thermostatic device with ethyl alcohol and soft cloth Clean the plastic parts and covers of the thermostatic device with dishwashing liquid water and soft cloth ...

Page 13: ...r your local MILKOTESTER dealer for further details MLKOTESTER service center will repair the product free of charge subject to terms condition mentioned below TERMS CONDITION 1 It covers only weighing balance purchased from authorized channel and does not covers accessories like Battery Adaptor RS232 cable 2 It does not cover the product which service manual and serial number has been altered rem...

Reviews: