3. USING BoneSTATION
This instrument is an in-vitro medical diagnostic device and it is usually installed in a
laboratory where samples and other biological tissues are present. For your safety, it
must therefore be cleaned and disinfected before use, and gloves must be worn when
using this instrument.
BoneSTATION has dedicated software, controlled by the LCD touchscreen terminal.
Heating power is auto-tuned by a PID controller. The operator must only set up the desired parameters
(Time and Temperature).
The device will always reproduce the same conditions of temperature in the set times, by automatically
adjusting the power level.
The following section shows how to run and modify the program values with BoneSTATION software. Use
the touch screen to select the desired menu.
3.1. Prepare the histomodule
Load the RACK with cassettes and position the CASSETTE FIXING DISC on top.
The fixing disc must always be used, in order to avoid possible spilling out of cassettes
from the rack during decalcification/fixation.
FIXING
DISC