en - Areas of application
45
Push the twin connector into the rinse attachment on the instru-
ment.
The third rinse aperture on the shaft of 5 mm rigid instruments does
not need to be connected for machine reprocessing.
When all instruments have been arranged, lower the rail to secure
the connectors and click it into place.
The water supply pipe retaining plates are secured in their positions
by the rail.
Water supply pipes which are not being used must also be positioned
under the rail so that they are not thrown about by the force of the
water during the cleaning process.
Instruments which cannot be secured by the retaining plates and
rail cannot be reprocessed in this module.
At the end
of a programme
Take the instruments out of the module by carrying out the process
steps in reverse.