Insert E 905/1 for short MIS instruments

,

Please read the Warning and Safety instructions in the operating Instructions for your G 7827/28 or PG 8527/28 Washer-disinfector carefully. You should only

process instruments/medical devices in this insert if they have been declared as suitable for machine processing by the manufacturer. Please also observe the
manufacturer's care instructions. The programme selected must be suitable for the materials from which the medical devices to be processed are made. Where
possible the programme should include thermal disinfection. The final rinse should, if possible, always be carried out with fully demineralised water.
If using automatic mobile unit recognition, please make sure that the coding is correctly set.

Miele Company Ltd., Fairacres, Marcham Road, Abingdon, Oxon, OX14 1TW, Tel 0845 365 0555,
Fax 0845 365 0777, Internet: http://www.miele.co.uk /

!

Alteration rights reserved / 3709

Items supplied:

Insert E 905/1, H 40, W 461, D 510 mm, fitted with:

16 connectors for adapters, injector sleeves and jets

2 notched rails for positioning the instruments

The insert is supplied with:

1 E 336 injector sleeve for MIS instruments, 121 mm long,

C

11 mm

2 x E 362 blanking screws

1 E 442 injector sleeve for MIS instruments, 121 mm long,

C

11 mm,

with cap

C

6 mm

2 x E 447 female adapters

]

for male Luer locks

\

4 x E 448 silicone tubes with screw thread, 300 mm long,

C

5 mm and Luer

Lock-Adapter

]

2 x E 449 male adapters

\

for female Luer locks

]

4 x E 452 jets, 2.5 x 60 mm, with screw thread

3 x E 453 jets, 4.0 x 110 mm, with retaining clip and screw thread

3 x E 454 jets for trocar sleeves, 10-15 mm, with screw thread

1 x E 464 holder for E 454 jet

1 x E 472 clamp for jets

C

4 mm

1 x E 907 insert/mesh tray with lid for small components

Areas of application:

Insert E 905/1 can be used in mobile units E 474/4, E 574/1, E 774/1 and E 902/1.
The mobile units are supplied with their own operating instructions.

The E 905/1 is designed for cleaning and disinfecting short MIS instruments, e.g.
those used for arthroscopy or laparascopy.
Instruments up to 190 mm long can be placed in the upper level, and up to 220 mm
long in the lower level.

Before loading the machine and before starting a programme, please check
that:

the silicone adapters in the injector manifold are intact and complete.

,

To ensure adequate water pressure, all positions must always be fitted with

jets or blanking screws.

For validation purposes please make sure that you follow the loading
instructions given on the template.

,

All MIS instruments must be dismantled for processing in accordance with

the manufacturer's instructions.

,

Long, narrow lumen instruments which cannot be taken apart must be

pre-washed in accordance with the manufacturer's instructions.

,

Before connecting a hollow instrument please make sure that all hollow

sections are free of blockages. Otherwise the cleaning solution will not be able
to access all areas properly.
Instruments should not be able to come into contact with one another during the
cleaning and disinfecting process.

,

Any hollow instruments that have become disconnected from the adapters

during processing must be reprocessed.

Injector sleeves

Injector sleeve E 442 is for processing hollow instruments and instruments with a
Ø less than 8 mm.

The diameter of the instruments should match that of the opening to the injector
sleeves/caps.

^

Insert trocar sleeves with a Ø less than 10 mm into the injector manifold. Connect
to the instrument's cleaning port with hose adapter E 448. Connect the hose
adapter to the injector manifold using a silicone adapter.

^

Insert hollow trocars into injector sleeves.

^

Connect to the instrument's cleaning port using an E 448 hose adapter, and
insert the hose adapter into a silicone adapter.

^

Insert long, narrow instruments which cannot be dismantled into an injector
sleeve. The functioning end must be open, if necessary using an E 472 spring
clamp to keep the handle open.
Connect to the instrument's cleaning port using an E 448 hose adapter. Connect
the hose adapter to the injector manifold using a silicone adapter.