1

Connect the BNC of the current probe to the oscilloscope (or other instruments), and connect 

the USB cable to power the current probe; 

2

Select  the  appropriate  range  on  the  probe  according  to  current  range,  the  corresponding 

button light will turn Green; 
3

Adjust the oscilloscope settings: input impedance 1MΩ;

 select probe to Current or display as A

Set attenuation factor for corresponding channel

range 100A

0.01V/A

  set to 100X

  range 10A

0.1V/A

set to 10X

 

4

Press Automatic zero adjustment button to realize zero adjustment of the probe

after success

the buzzer 

beep

” one time

i

f “

beep

” three times

indicating that the zero adjustment has failed

Manual zero adjustment is also possible

the external magnetic field may have a slight influence 

on the DC zero position of the probe

do not move it in a large range after the zero adjustment is 

completed; 

5

Open the jaws to clamp the conductor under test according to the direction indicated by the 

clamp head. Note: If the measured current flows in the opposite direction to the clamp head, the 
output will be negative; 

6

Adjust the oscilloscope to get the best waveform; 

 
Note: When the current exceeds the range, the buzzer will beep for a long time and the button 
light will flash. 

5.

 

Maintenance 

During the warranty period of the product (one-year) and under normal use, the company will be 
responsible  for  free  repairs  due  to  fault  caused  by  the  quality  of  the  product  itself,  and  the 
product must not been disassembled and repaired without Micsig permission. 
Please keep the product dry, clean and tidy. If there is dirt, use a soft cloth or sponge with alcohol 
to wipe off the dirt. Do not use water. 
In order to ensure the performance of the product, it is recommended to check or calibrate once 
a year. 
 
 
 
 
 
 
 
 
 

Summary of Contents for CP2100A

Page 1: ...CP2100A CP2100B ...

Page 2: ...l and any information contained in it including but not limited to implied guarantees for merchantability and suitability for specific purposes Micsig is not responsible for errors or accidental or consequential damages caused by the provision use or use of this document or any information contained in it If a separate written agreement has been reached between Micsig and the user that contains wa...

Page 3: ...3 10 1 Product Introduction 5 2 Appearance 5 3 Specifications 6 4 Instructions 9 5 Maintenance 9 ...

Page 4: ...touch the measured conductor and sensor head during measurement Do not use in a humid environment Do not touch the instrument or the measured object with wet hands Please use this product within the range allowed by the end user Please use this product as required Please ground this product through the USB power cord ...

Page 5: ...ters via connector The maximum current can be measured to 100Apk 70Arms it has 2 models CP2100A and CP2100B CP2100A can measure bandwidth to DC 800KHz CP2100B can measure bandwidth to DC 2MHz 2 optional measuring ranges 10A and 100A Automatic and manual zero adjustment powered by USB no additional power supply making measurement more convenient Often used in motor drives industrial frequency inver...

Page 6: ...k 70 7Arms DC ACpk 200Apk pk 70 7Arms AC Maximum working voltage CAT III 300V CAT II 600V Maximum floating voltage CAT III 300V CAT II 600V Maximum conductor diameter 13mm Overload indication The buzzer sounds long and the button light flashes Supplied power DC 5V Probe head size 11 6 1 2 5cm Control box size 10 8 5 6 2 6cm Length 228cm Weight 290g Package weight 1000g Package dimensions 29 5 23 6...

Page 7: ...7 10 Maximum current VS Frequency curve Amplitude frequency characteristic curve CP2100A ...

Page 8: ...8 10 Amplitude frequency characteristic curve CP2100B DC signal linearity 0 01V A ...

Page 9: ...sition of the probe do not move it in a large range after the zero adjustment is completed 5 Open the jaws to clamp the conductor under test according to the direction indicated by the clamp head Note If the measured current flows in the opposite direction to the clamp head the output will be negative 6 Adjust the oscilloscope to get the best waveform Note When the current exceeds the range the bu...

Page 10: ...10 10 Shenzhen Micsig Instruments Co Ltd Tel 86 755 88600880 Email sales micsig com http www micsig com ...

Reviews: