<Command>
*IDN?
Function
:
It queries the device basic data.
Description
:
Return dataMICROTEST,6632-30G,0,1.194<field1>,<field2>,<field3>,<field4>
<field1>manufacturer
<field2>model number
<field3>serial number or 0
<field4>firmware revision
MICROTEST,6632-30G,0,1.194
<Command>
*OPC
Function
:
It sets Standard Event Status Register to 0 bit,when all operations are completed.
<Command>
*OPC?
Function
:
When all operations are completed, 1 will be returned.
<Command>
*OPT?
Function
:
It queries the installed in the system interface.
Description
:
If 6632 MAX test frequency is 1MHz , it returnsF01
If 6632 MAX test frequency is 3MHz, it returns F03
If 6632 MAX test frequency is 5MHz, it returns F05
If 6632 MAX test frequency is 10MHz, it returns F10
If 6632 MAX test frequency is 15MHz, it returns F15
If 6632 MAX test frequency is 20MHz, it returns F20
If 6632 MAX test frequency is 30MHz, it returns F30
<Command>
*RST
Function
:
It aborts all pending operations, and sets 6632 to its initial setups.
Description
:
Initial Setups
Parameters:Ls,Q,Z,
θdeg
FREQuency:1kHz
LEVEL:1Vac
SPEED:MED.
TRIGger:REPEAT
<Command>
*SRE <NR1>
*SRE?
Function
:
It sets or queries the status of Service Request Enable Register.
Description
:
Set parameter 0~255
Set syntax*SRE 1
Query syntax*SRE?
Return data 0~255 (Format is in <NR1>)
<Command>
*STB?
Function
:
It queries the status of Service Request Status Register
Description
:
Return data<NR1>
Summary of Contents for 6632 Series