- 43 -
Appendix
Common Commands
Command
Function
Description
*IDN?
Identification query
Returns the data identifying the instrument.
(e.g. the data output will be:
‘MICROTEST,6630E-10,1.000’ where the first
field is the manufacturer, then the model
number, then a zero and the software revision
number: here represented as Issue 1.0).
*TRG
Trigger and query
Triggers a direct measurement and returns the
results to the controller.
:TRIG
Trigger
Triggers a direct measurement, but does not
return the results to the controller.
:TRIG?
Trigger and return
Triggers a direct measurement and returns the
results to the controller. This is as same
as ’*TRG’ command.
:DISP:PAGE
<MEAS|SWEEP|CAL|SYSTE
M}
Select measuring mode
MEAS Meter mode
SWEEP Sweep mode
CAL Calibration mode
SYSTEM System mode
:DISP:PAGE?
Returns the measurement mode of the
currently selected step.
:MEAS:PARA?
Returns the parameter of the currently selected
step.
:MEAS:FREQ
<frequency
NR3>
Set frequency of AC measurement
Frequency range : 10.0 ~ 30000000.0 Hz
:MEAS:FREQ?
Frequency query
:MEAS:VOLT:AC <voltage
NR3>
Set voltage of AC measurement
Voltage range : 0.01-2.0 V
:MEAS:VOLT:AC?
AC voltage query
:MEAS:CURR:AC <current
NR3>
Set current of AC measurement
Current range
Output Resistance 100 Ohm : 0.0001-0.02 A
Output Resistance 25 Ohm : 0.0002-0.04 A
:MEAS:CURR:AC?
AC current query
:MEAS:VOLT:DC <voltage
NR3>
Set voltage of DC measurement
Voltage range : 0.01-2.0 V