Quick Start Guide 

 

 
 
MIP Soft & Hard Iron Calibration  

 

3DM-GX3-25 

(fw version 2.0.00 and above)

 

 

3DM-GX3-35 

 

3DM-GX3-45 

 

 
Overview 

Most models of the 3DM-GX3

®

 series contain a magnetometer.  The magnetometer values are 

available as fully calibrated outputs and are also used to correct heading drift in the Orientation 
Matrix, Euler Angles and Quaternion outputs.  The 3DM-GX3

®

 has been carefully designed to 

eliminate as many ferro-magnetic components as possible to prevent distortions in the readings 
of the earth’s magnetic field which can cause errors in the magnetometer output.  In addition, the 
magnetometer is factory calibrated on non-magnetic stages in order to compensate for any 
remaining internal sources of error.  However, these measures can only compensate for error 
sources that are internal to the device; they cannot compensate for errors that may be introduced 
externally by mounting structures or adjacent devices.  Careful mounting of the 3DM-GX3

®

 can 

avoid external magnetic anomalies caused by sources such as coils, magnets, and ferrous metal 
structures and mounting components (steel nuts and bolts).  Often these sources are hard to avoid 
or are hidden and so a field calibration of the magnetometer after final installation is highly 
recommended.  This can be accomplished using MicroStrain’s 

MIP Soft & Hard Iron 

Calibration software

 

Important 

If you have a 3DM-GX3

®

-25 with a firmware version that is lower than 2.0.00, you need to use 

the 3DM-GX3

®

 Soft & Hard Iron Calibration.  If you have version 2.0.00 or above, you may use 

either 3DM-GX3

®

 Soft & Hard Iron Calibration or the MIP Soft & Hard Iron Calibration 

software. 

Caution 

Starting with 3DM-GX3

®

 firmware version 1.1.27, the user has the option of turning off the 

heading correction and/or the magnetometer.  The default setting is to have both turned on. 
Please make sure the magnetometer is turned 

ON

 before calibrating. 

 

Instructions

  

Before starting, make sure you have installed the MIP Soft & Hard Iron Calibration software 
from the installation CD or from the installer downloadable a

www.microstrain.com

. 

 

1)

 

Connect the 3DM-GX3

®

 to your computer using either the RS-232 communication cable 

and power supply or the USB communication cable.  If you have not already done so, 
follow the instructions in the 3DM-GX3

®

 Quick Start Guide on the installation CD to 

make sure the device is properly installed and communicating with your computer. 

Summary of Contents for 3DM-GX3 Series

Page 1: ...ts and bolts Often these sources are hard to avoid or are hidden and so a field calibration of the magnetometer after final installation is highly recommended This can be accomplished using MicroStrai...

Page 2: ...appear on the display in 3D c Click Stop Collecting d Click Spherical Fit and or Ellipsoid Fit to show the calculated calibration result e Click Corrected Fit to show how the corrected data will look...

Page 3: ...resenting the center of the data sphere is offset from the circle which represents the center of the sensor Normally these should be superimposed Figure 3 Spherical Fit The sphere fits pretty well but...

Page 4: ...ure 4 Ellipsoidal Fit The ellipsoid fit captures many of the points that were not on the surface of the spherical fit Figure 5 Corrected Fit The green sphere shows what the calibrated data should look...

Page 5: ...en points represent data taken after the calibration is written to the device Notice the values in the Device Iron Cal Matrix and Offset Figure 7 Live Data and Corrected Fit This shows how the correct...

Page 6: ...f your application Mounting changes or changes in surroundings may make a difference in the magnetometer output and by using this application you can see the magnetic effects and correct for them as n...

Page 7: ...use permanent damage to its components Many GPS antennas have a magnetic base the antenna we provide does not If you substitute a GPS antenna and its base is magnetic please keep it away from the 3DM...

Reviews: