11
connect the wires with terminals on the plug core correctly (the
connection terminals are signed with clear numbers). Please
choose Φ4.5mm~7mm shield cable and connect the cable reliably
to prevent short circuit.
d) Pull the cable slightly and push the plug core into housing (a
rattling sound could be listened), then screw down the cable-fixed
nut.
e) To remove the cable, loose the cable-fixed nut to relax the cable,
and operate as Item2 to disconnect the plug. Take out the cable
from terminals with little screwdriver and pull out the cable from
cable-fixed nut, then renew the connection between plug core,
plug and socket.
Note:
renew the rectangle-ring to the pre-mounting situation in order
tomaintain the protection class when connect plug and socket.
4.4 Indicator Operation
(
Only the header with a transmitter
)
4.4.1 Indicator Adjustment
a) Open the front panel of indicator as Fig.8(take LED indicator as
example, LCD indicator is similar).
b) Input 4mA
(
±0.05%
)
into indicator, adjust zero potentiometer W2
and show “000”;
c) Then input 20mA
(
±0.05%
)
into indicator, adjust FS potentiometer
W1 and show the required value. The adjustment ends.
Note:
the indicator is 2-wire, and should be connected serially with
Summary of Contents for MPM480
Page 1: ...MPM480 Operation Manual ...
Page 20: ......