Step 2 — Connect the System

Caution

: Be certain all cables are connected BEFORE applying 

power. Always power down BEFORE disconnecting cables.

To connect the system, do the following:

1. Connect the reader (

1

) to 

the “SCANNER” connector 

on the IC-331 interface 

converter (

2

) using the 

attached 3-foot cable.

1

2. Connect the IC-331 

directly to the “SCANNER” 

15 pin connector on the 

IB-131 (

3

).

3. Connect the Host Com-

puter (

4

) to the IB-131 

“HOST” 25-pin connector 

via the Null Modem Con-

figuration Cable (

5

).

2

 

4. Connect Power Supply 

(

6

) to the IB-131 

“POWER” connector.

5. Apply power to the system.

1. Since power supply is included in the single cabling assembly, the reader cable 

should not exceed 3 feet. RS-232 cabling from the IB-131 to the host can be up 

to 47 feet provided it does not include power input

2. If using your own null modem RS-232 host cable, be certain that the host’s TxD 

connects to the reader’s RxD and the reader’s TxD connects to the host’s RxD.

Hardware Connections

4

1

2

3

7

5

Scanner

Ne

tw

or

k

Ho

st

6

Summary of Contents for MS-3 CCD

Page 1: ...8 or above you can use ESP Easy Setup Program to configure and control your MS 3 CCD Reader In addition to ESP you can send commands to your reader by serial command For a full explanation of the MS 3...

Page 2: ...s If connecting to a host with an IB 131 Interface Box you will need the following Itema a See Connect the System on page 3 for diagram of system Part Number Description 1 FIS 0003 1XXXG MS 3 CCD Read...

Page 3: ...nnect the Host Com puter 4 to the IB 131 HOST 25 pin connector via the Null Modem Con figuration Cable 5 2 4 Connect Power Supply 6 to the IB 131 POWER connector 5 Apply power to the system 1 Since po...

Page 4: ...ware from the main menu 3 Launch Setup exe under ESP and follow the prompts If downloading from the web 1 Go to http www microscan com downloadcenter 2 Create a new myMicroscan member account or if yo...

Page 5: ...click OK If you do not want to make this selection every time you load ESP uncheck Show this window at Startup 2 Select the default name MS 3 CCD 1 for example or type in a file name of your choice an...

Page 6: ...the Start button When connected the reader s settings will be loaded into ESP and the CONNECTED message will appear in a green box in the status bar at the bottom right of your screen 3 If the connect...

Page 7: ...bol type default is 10 and 6 2 Avoid bright light or IR light from other sources including other readers 3 Pitch the symbol or reader at a minimum of 15 to avoid specular reflection the return of dire...

Page 8: ...ode 1 Click the Test button in Setup Mode to begin the read rate test 2 Follow the instructions on the Setup Mode screen 3 To end the read rate test click the Stop button Note If the reader is in the...

Page 9: ...pp Mode button To return to Setup Mode click the Setup Mode button By Serial Command From your terminal program or the Terminal screen in ESP you can enter serial commands and configuration and utilit...

Page 10: ...10 MS 3 CCD Reader Quick Start Guide...

Reviews: