GUIDANCE AND MANUFACTURER'S
DECLARATION: ELECTROMAGNETIC EMISSIONS
The Spider SAS is intended for use in the electromagnetic environment specified below. The
customer or the user of the Spider SAS should ensure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
Electromagnetic radiation
disturbance (radiated
emissions) CISPR 11
Group 1
The Spider SAS uses RF energy only for its
internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference to nearby electronic equipment.
Disturbing voltage at
supply terminals
conducted emissions
CISPR 11
Class B
The Spider SAS is suitable for use in a
professional health care establishment and
home health care environment.
Harmonic current
emissions IEC 61000-3-2
Non applicable
Voltage variations, voltage
fluctuations and flicker
IEC 61000-3-3
Non applicable
GUIDANCE AND MANUFACTURER'S DECLARATION: MAGNETIC
AND ELECTROMAGNETIC IMMUNITY
Spider SAS is intended for use in the electromagnetic environment specified below. The
customer or the user of the Spider SAS should ensure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic environment
- guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±8 kV
contact
±15 kV air
± 8 kV
contact
± 15 kV air
Professional health care
establishment and home health care
environment.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
Non
applicable
Professional health care
establishment and home health care
environment.
± 1 kV for
input/output lines
Surge
IEC 61000-4-5
± 1 kV
between phases
Non
applicable
Professional health care
establishment.
± 2 kV
between line(s) to
earth
Magnetic field
at industrial
rated
frequency
(IEC61000-4-8)
30 A/m
Non
applicable
Professional health care
establishment.
16.3.
16.4.
16. APPENDIX 2: ELECTROMAGNETIC COMPATIBILITY
SPIDER SAS – PHYSIO HUB – HOOKUP RS – U908F