![Micronor MR380-1 Series Instruction Manual Download Page 11](http://html1.mh-extra.com/html/micronor/mr380-1-series/mr380-1-series_instruction-manual_1788067011.webp)
MR380-1 Series Instruction Manual
Page 11 of 35
3.2.
Mounting the Controller Module
The controller unit mounts on DIN rails as well as a single mounting clip. Both mounting schemes
Figure 7. Mounting Options: DIN Rail Mount versus Chassis Mount
3.3.
Connecting the Controller
An LC Duplex optical cable is used to interconnect the sensor and controller. The sensor
incorporates a 3m optical pigtail (or alternate length as specified by customer). If a longer
connection to the controller is required, then an extension cable having LC duplex connector may
be used.
Remove LC Duplex connector's dust cap(s) and open the external shutter of the controller. When
inserting the LC Duplex connector, there should be a positive click when the connector is engaged
properly. Avoid repeated fiber optic disconnects to reduce likelihood of surface contamination.