Micronix MP-21 Series Reference Manual Download Page 7

MP-

21

 

Linear Actuator

 

 

 
 

 

Rev: 3.07 

5  

Reference Manual 

MICRONIX USA,LLC 

 

www.micronixusa.com 

 
 

3.

 

Preparing to Install the MP-21 

  

3.1

 

Installation Preparation 

 

The  actuator  is  calibrated  and  guaranteed  to  be  within  specification  at  20°C 
±5°C. Be sure to use the actuator under the following conditions: 
 

 

Mount to a clean and flat surface which is free of debris, burrs or dings 

 

An  indoor  atmosphere  free  of  corrosive  gases,  excessive  dust,  and 
condensation 

 

Temperature range of 0-40°C 

 

Relative humidity between 20-80% 

 

Locate away from water, heat, and electrical noise 

 
 

3.2

 

Package Contents 

 

 

If product is damaged or there are missing components, contact MICRONIX USA 
immediately. Do not discard product packaging in case of return shipment.  
 

Package Should Contain: 

 

MP-21 Linear Actuator

 

 

Locking Nut & Wrench

 

 

Reference Manual

 

 

Any other previously agreed upon components such as a controller and 
cable

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Summary of Contents for MP-21 Series

Page 1: ...S e r i e s Reference Manual Linear Actuator MP 21...

Page 2: ...MP 21 Linear Actuator Reference Manual Rev 3 07 MICRONIX USA LLC Tel 949 480 0538 Fax 949 480 0538 Email info micronixusa com http micronixusa com...

Page 3: ...Installation 6 5 Connecting the MP 21 Linear Actuator 7 5 1 Atmospheric Environments 7 5 2 Vacuum Environments 7 5 2 1 Handling and Preparation 7 5 2 2 Installation Wiring Diagram 7 6 Dimensions 9 6 1...

Page 4: ...s achieved through use of a stepper motor and limit switch Aluminum construction provides for a rigid housing while reducing weight The MP 21 is ideally suited for motorization of micro meter replacem...

Page 5: ...Data Motor SM 001 Pitch mm 0 5 Speed Max mm sec 5 Resolution Calculated m 2 5 Full Step Resolution Typical m 0 1 Bi directional Repeatability m 0 5 Uni directional Repeatability m 0 4 1 4 Load Charac...

Page 6: ...USA LLC www micronixusa com 2 Model Configurations 2 1 MP 21 Order Numbers Contact MICRONIX USA for custom versions and stacking configurations Order No MP 21 1 0 1 SM 001 1 12 mm 1 25 mm 2 Optional E...

Page 7: ...of debris burrs or dings An indoor atmosphere free of corrosive gases excessive dust and condensation Temperature range of 0 40 C Relative humidity between 20 80 Locate away from water heat and elect...

Page 8: ...o the housing The MP 21 is designed for axial loading do not apply radial loads to the plunger When preparing the stage for vacuum environments take the necessary precautions such as wearing gloves cl...

Page 9: ...g gloves clean room clothing etc when handling the precision mechanical device as to avoid any contaminants Maximum Bake out temperature is 100 C MICRONIX USA can supply the actuator with a vacuum com...

Page 10: ...MP 21 Linear Actuator Rev 3 07 8 Reference Manual MICRONIX USA LLC www micronixusa com Temporary test connectors are standard Requires customer supplied feed through connector...

Page 11: ...MP 21 Linear Actuator Rev 3 07 9 Reference Manual MICRONIX USA LLC www micronixusa com 6 Dimensions 6 1 MP 21 12 mm 6 2 MP 21 25 mm Weight 0 14 Kg Weight 0 17 Kg...

Page 12: ...in motion to avoid personal injury Follow the Safe Operating Environment requirements and use proper cable management to ensure a clean and safe operating environment The MP 21 is designed to support...

Page 13: ...H Pitch 0 5 mm rev Resolution Full step 2 5 m Wire Color Pin Function Atmospheric 1 Motor A Orange 2 Motor A White 3 Motor B Yellow 4 Motor B Black 5 Limit Switch GND Green 6 Limit Switch Blue 7 Limit...

Page 14: ...tches in the controller will result in physical damage to the MP 21 Linear Actuator Travel Length mm Software Limit Switch Position mm 12 12 or less 25 25 or less The mechanical limit switch is factor...

Page 15: ...ub 9 Pin 1 5m Wiring Diagram A 3 1 Straight Through 9 Pin Feed Through Connector Pinout Description Color L1 L2 L3 L4 L5 Motor Phase A Green 5 5 1 1 Green 1 Motor Phase A Green White 4 4 2 2 White Gre...