P
OWER
The Smart VacPrep is designed to operate with line voltage of 100/120/230/240VAC ±10, 50/60
Hz through a standard wall receptacle. Noise-free power of the correct voltage and frequency,
with a safety earth ground, should be available through a standard wall receptacle. There should
be a minimum 15A rated breaker @ 100/120 VAC and a minimum 7.5A @ 240 VAC.
The analyzer and peripheral devices
must
be installed on their own dedicated power
line. Other devices — such as motors, generators, or ovens —
should not
be placed
on the same power line.
Replacement power supply cords must be rated for the specifications stated above.
I
NSTALL THE
V
ACUUM
P
UMP
The
Vacuum Pump Guide
contains installation instructions for vacuum pumps and can be found
on the Micromeritics web page (
).
Power
Smart VacPrep Installation Instructions and Checklist
067-42870-02 (Rev A) — June 2022
1 - 3