Performing an Analysis
ASAP 2460 Operator’s Manual
2-42
246-42800-01 - Aug 2013
Krypton Analysis
Use the following guidelines for performing krypton analyses:
• All ports must be idle in order to start an analysis.
• All analyses must use krypton gas.
• From one to five analyses can be started at the same time. The last port is used to store Krypton for
dosing (Port 2, 4, or 6 depending on your system configuration).
• An empty sample tube must be installed in the last port (2, 4, or 6 depending on your system
configuration).
• The sample file used for an analysis must specify the Dosing Method:
From last port
in the
Adsorptive Properties.
• Samples cannot be removed from or added to ports until the full set of analyses has
completed
The data collection is done sequentially — one analysis starts and completes before the next is started.
1.
Install an empty sample tube on the port to be used to store krypton (the last port: 2, 4, or 6) and
place a dewar cover over the sample tube stem and the p
o
) tube stem.
Krypton analyses are available only if the Krypton option is installed on the analyzer.
The dewar below the port used to store and purify krypton must be at least 50% full
of the analysis bath liquid and an empty sample tube must be installed on the port
before starting a krypton analysis.
Summary of Contents for ASAP 2460
Page 4: ......
Page 12: ...Table of Contents ASAP 2460 Operator s Manual viii 246 42800 01 Nov 2013...
Page 30: ...Using the Software ASAP 2460 Operator s Manual 1 18 246 42800 01 Aug 2013...
Page 93: ...ASAP 2460 Operator s Manual Generating Graph Overlays 246 42800 01 Aug 2013 2 63...
Page 94: ...Generating Graph Overlays ASAP 2460 Operator s Manual 2 64 246 42800 01 Aug 2013...
Page 202: ...Service Test ASAP 2460 Operator s Manual 4 36 246 42800 01 Aug 2013...
Page 230: ...Report Examples 2460 Operator s Manual 5 28 246 42800 01 Aug 2013 Isotherm Linear Plot...
Page 231: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 29 BET Surface Area Report...
Page 232: ...Report Examples 2460 Operator s Manual 5 30 246 42800 01 Aug 2013 BET Surface Area Plot...
Page 233: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 31 t Plot Report...
Page 236: ...Report Examples 2460 Operator s Manual 5 34 246 42800 01 Aug 2013...
Page 242: ...Service Test Mode 2460 Operator s Manual 6 6 246 42800 01 Aug 2013...
Page 270: ...Calibrating the System ASAP 2460 Operator s Manual 7 28 246 42800 01 Aug 2013...
Page 276: ...Ordering Information ASAP 2460 Operator s Manual 8 6 246 42800 01 Aug 2013...
Page 278: ......
Page 280: ......
Page 374: ...Appendix C ASAP 2460 Operator s Manual C 50 246 42800 01 Aug 2013...
Page 398: ...Appendix F ASAP 2460 Operator s Manual F 18 246 42800 01 Aug 2013...
Page 412: ...Index ASAP 2460 Operator s Manual Index 8 246 42800 01 Aug 2013...
Page 413: ...ASAP 2460 Operator s Manual Appendix G Aug 2013 SVP 1 SVP Smart VacPrep Operator Manual...
Page 414: ...Appendix G ASAP 2460 Operator s Manual SVP 2 Aug 2013...
Page 422: ...Contact Us vi Smart VacPrep Operator Manual...
Page 432: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 466: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...