Appendix D
ASAP 2460 Operator’s Manual
D-2
246-42800-01 - Aug 2013
• Helium is unavailable; free space determination by the application software requires the use of
helium
• Analysis speed is a major factor; a helium free space measurement of 10 to 15 minutes is required
• Your sample tends to absorb and retain helium for a prolonged period of time or if it adsorbs helium
Calculate
This method is the most rapid and efficient way of compensating for free space. You must ensure the
following is accomplished:
• Perform a blank analysis on the sample tube
• Load the blank analysis file data into the sample tube file
• Enter the analysis bath temperature (found on the p
o
and Temperature dialog)
• Enter the sample mass and density (found on the Sample Information dialog)
Enter
This method allows you to enter predetermined values for the warm and cold free spaces. The values to
enter may be obtained in one of two ways:
• A pre-analysis free space calibration of the sample tube containing sample
• The total free space of an empty sample tube is measured and the displacement of the sample
calculated from its mass and density and subtracted from the total free space
In either procedure, ensure that the level (or, in cases where the Isothermal Jacket is used, the effective
level) of the cryogen bath on the sample tube is the same when the analysis is performed as it was
when gathering data for free space calculations.
Summary of Contents for ASAP 2460
Page 4: ......
Page 12: ...Table of Contents ASAP 2460 Operator s Manual viii 246 42800 01 Nov 2013...
Page 30: ...Using the Software ASAP 2460 Operator s Manual 1 18 246 42800 01 Aug 2013...
Page 93: ...ASAP 2460 Operator s Manual Generating Graph Overlays 246 42800 01 Aug 2013 2 63...
Page 94: ...Generating Graph Overlays ASAP 2460 Operator s Manual 2 64 246 42800 01 Aug 2013...
Page 202: ...Service Test ASAP 2460 Operator s Manual 4 36 246 42800 01 Aug 2013...
Page 230: ...Report Examples 2460 Operator s Manual 5 28 246 42800 01 Aug 2013 Isotherm Linear Plot...
Page 231: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 29 BET Surface Area Report...
Page 232: ...Report Examples 2460 Operator s Manual 5 30 246 42800 01 Aug 2013 BET Surface Area Plot...
Page 233: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 31 t Plot Report...
Page 236: ...Report Examples 2460 Operator s Manual 5 34 246 42800 01 Aug 2013...
Page 242: ...Service Test Mode 2460 Operator s Manual 6 6 246 42800 01 Aug 2013...
Page 270: ...Calibrating the System ASAP 2460 Operator s Manual 7 28 246 42800 01 Aug 2013...
Page 276: ...Ordering Information ASAP 2460 Operator s Manual 8 6 246 42800 01 Aug 2013...
Page 278: ......
Page 280: ......
Page 374: ...Appendix C ASAP 2460 Operator s Manual C 50 246 42800 01 Aug 2013...
Page 398: ...Appendix F ASAP 2460 Operator s Manual F 18 246 42800 01 Aug 2013...
Page 412: ...Index ASAP 2460 Operator s Manual Index 8 246 42800 01 Aug 2013...
Page 413: ...ASAP 2460 Operator s Manual Appendix G Aug 2013 SVP 1 SVP Smart VacPrep Operator Manual...
Page 414: ...Appendix G ASAP 2460 Operator s Manual SVP 2 Aug 2013...
Page 422: ...Contact Us vi Smart VacPrep Operator Manual...
Page 432: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 466: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...