ASAP 2460 Operator’s Manual
Analysis Conditions Files
246-42800-01 - Aug 2013
3-19
Target Pressures
Options
button
•
First Pressure Fixed Dose
-
Select this option to have the sample dosed with a specified amount of
gas until the first pressure point is reached. This initial dosing quickly
meets the adsorptive demand of the sample.
The first point on the pressure table is the threshold value. Once this first
pressure point is reached, points are equilibrated and recorded in
accordance with the specified pressure table.
Use this mode if performing a standard nitrogen analysis of microporous
or highly mesoporous materials such as catalysts. If the first pressure
table point is low, and the gas uptake of the sample is expected to be
high, choosing this mode can shorten the time required to reach the first
point on the pressure table.
•
Maximum volume increment
-
Select this option to specify when additional data points are to be
collected between target pressures in regions of high adsorption. When
the maximum increment has been adsorbed since the last collected data
point, another point is equilibrated and collected.
When using this option, reaching pressure points exactly is not
important; therefore, the tolerances should be set relatively large
(10 mmHg and 10% or so) for proper functioning of the algorithm. The
pressure table should also have several points scattered over the region
of interest.
During desorption, this field is treated as a maximum volume
“decrement” value.
Field or Button
Description (continued)
Summary of Contents for ASAP 2460
Page 4: ......
Page 12: ...Table of Contents ASAP 2460 Operator s Manual viii 246 42800 01 Nov 2013...
Page 30: ...Using the Software ASAP 2460 Operator s Manual 1 18 246 42800 01 Aug 2013...
Page 93: ...ASAP 2460 Operator s Manual Generating Graph Overlays 246 42800 01 Aug 2013 2 63...
Page 94: ...Generating Graph Overlays ASAP 2460 Operator s Manual 2 64 246 42800 01 Aug 2013...
Page 202: ...Service Test ASAP 2460 Operator s Manual 4 36 246 42800 01 Aug 2013...
Page 230: ...Report Examples 2460 Operator s Manual 5 28 246 42800 01 Aug 2013 Isotherm Linear Plot...
Page 231: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 29 BET Surface Area Report...
Page 232: ...Report Examples 2460 Operator s Manual 5 30 246 42800 01 Aug 2013 BET Surface Area Plot...
Page 233: ...2460 Operator s Manual Report Examples 246 42800 01 Aug 2013 5 31 t Plot Report...
Page 236: ...Report Examples 2460 Operator s Manual 5 34 246 42800 01 Aug 2013...
Page 242: ...Service Test Mode 2460 Operator s Manual 6 6 246 42800 01 Aug 2013...
Page 270: ...Calibrating the System ASAP 2460 Operator s Manual 7 28 246 42800 01 Aug 2013...
Page 276: ...Ordering Information ASAP 2460 Operator s Manual 8 6 246 42800 01 Aug 2013...
Page 278: ......
Page 280: ......
Page 374: ...Appendix C ASAP 2460 Operator s Manual C 50 246 42800 01 Aug 2013...
Page 398: ...Appendix F ASAP 2460 Operator s Manual F 18 246 42800 01 Aug 2013...
Page 412: ...Index ASAP 2460 Operator s Manual Index 8 246 42800 01 Aug 2013...
Page 413: ...ASAP 2460 Operator s Manual Appendix G Aug 2013 SVP 1 SVP Smart VacPrep Operator Manual...
Page 414: ...Appendix G ASAP 2460 Operator s Manual SVP 2 Aug 2013...
Page 422: ...Contact Us vi Smart VacPrep Operator Manual...
Page 432: ...2 About the Software 2 4 Smart VacPrep Operator Manual...
Page 466: ...8 Error Messages 8 4 Smart VacPrep Operator Manual...