AquaPrep 055 

Organization of the Manual

Dec 01                                                  

1-1 

1.  GENERAL INFORMATION

Organization of the Manual

This manual describes the installation and operation of the AquaPrep 055; its contents are 
organized as follows:

Conventions

This manual uses the symbols shown below to identify notes of importance, warnings, and 
cautions.

Chapter 1

GENERAL INFORMATION

 

Provides a general description of the AquaPrep, safety precautions, 
and specifications.

Chapter 2

INSTALLATION

Describes how to unpack, inspect, and install the AquaPrep.

Chapter 3

OPERATION

Provides instructions for operating the AquaPrep.

Chapter 4

TROUBLESHOOTING

Provides troubleshooting information.

Chapter 5

ORDERING INFORMATION

Provides ordering information for AquaPrep accessories.

Notes contain important information pertinent to the subject matter.

Warnings contain information that help you prevent actions that may cause personal 
injury.

Cautions contain information that help you prevent actions that may damage the 
analyzer.

Summary of Contents for AquaPrep 055

Page 1: ...AquaPrepTM 055 Operator s Manual 055 42801 01 October 2009...

Page 2: ...Micromeritics and Saturn DigiSizer are trademarks of Micromeritics Instrument Corporation Teflon is a registered trademark of Norton Company Micromeritics Instrument Corporation 2002 2009 All rights...

Page 3: ...e following conditions 1 Certain products sold by MICROMERITICS are the products of reputable manufacturers sold under their respective brand names or trade names We therefore make no express or impli...

Page 4: ......

Page 5: ...Description 1 3 Specifications 1 4 2 INSTALLATION Unpacking and Inspecting the Equipment 2 1 Equipment Damage or Loss During Shipment 2 1 Equipment Return 2 1 Installing the AquaPrep 2 2 Selecting the...

Page 6: ......

Page 7: ...uaPrep safety precautions and specifications Chapter 2 INSTALLATION Describes how to unpack inspect and install the AquaPrep Chapter 3 OPERATION Provides instructions for operating the AquaPrep Chapte...

Page 8: ...sis Tap water varies greatly in its dissolved air content and the presence of debris This water must be deionized and well filtered before using with the AquaPrep Filtering the water through at least...

Page 9: ...and outlet connectors on its front panel The switch for turning it on and off is located on the rear panel just left of the power card entrance Figure 1 1 The AquaPrep 055 The air removed from the wat...

Page 10: ...Accommodates to standard power mains worldwide Voltage 100 120 220 240 VAC 10 Current 2 0 A 100 120 VAC 1 0 A 220 240 VAC Frequency 50 60 Hz Water Filtered and deionized EXPOSED MATERIALS Polypropyle...

Page 11: ...rrier of the equipment immediately Equipment Return Micromeritics strives to ensure that all items arrive safely and in working order Occasionally due to circumstances beyond our control equipment is...

Page 12: ...truments leave the factory set for 120 VAC and with the line fuse removed The correct setting of the universal power entrance must be checked and the appropriate fuse installed before the AquaPrep can...

Page 13: ...t out of the power connector housing 4 Orient the voltage selector card so that the desired voltage is indicated at the bottom Orient the indicator pin so that it points upward as shown in the followi...

Page 14: ...e side with the double fuse slots is away from the cover b If the fuse block is positioned properly for the desired fusing proceed to Step c If the fuse block is not positioned properly 1 Remove the f...

Page 15: ...put power connector and snap it into place Once the fuse block and cover assembly are in place the position of the indicator pin shows the input power selected The fuses used in the AquaPrep must be i...

Page 16: ...t panel Connecting the Power Cord Insert one end of the power cord into the input power connector on the rear panel of the Aqua Prep and the other end into an appropriate power source Do not turn the...

Page 17: ...m or 2 in the surface of the water so that splashing does not re introduce air into the liquid It is important that the tubing is arranged at different levels This promotes circulation and ensures tha...

Page 18: ...dequacy of deaeration For example you may observe 20 bubbles per minute If the background measurement you will perform next shows adequate deaeration you can watch for approximately 20 bubbles per min...

Page 19: ...for ordering information Hydrophobic Capsule Replace the hydrophobic capsule when excessive deterioration is detected Refer to Chapter 5 for ordering information 1 Unplug the AquaPrep from its power...

Page 20: ......

Page 21: ...ccessories can be ordered by telephone 770 662 3636 email customerservice micromeritics com internet www micromeritics com Part Number Item and Description 003 51130 01 Fuse 1 0 Amp 5 x 20 mm slow blo...

Page 22: ......

Reviews: