AquaPrep 055
Installing the AquaPrep
Oct 09
2-5
c Insert appropriate fuse(s) for the input power source. Refer to the chart below for the
appropriate fuse rating.
7. Insert fuse block and cover assembly into input power connector and snap it into place.
Once the fuse block and cover assembly are in place, the position of the indicator pin
shows the input power selected.
The fuses used in the AquaPrep must be identical in type and rating to that specified.
Use of other fuses could result in electrical shock and/or damage to the analyzer.
Power Source
Fuse
100-120 VAC
2.0 Amp (slow-blow)
220-240 VAC
1.0 Amp (slow-blow)