10. Using an appropriate heating station adapter, insert the sample tube into the heating
station.
11. Allow the sample(s) to degas for the desired length of time. One to two hours is typical for
many samples at temperatures above 200 ºC. To determine the optimum time, select the
highest safe temperature for your sample, then degas it for various intervals, determining
the surface area after each interval.
For example, degas the sample for 15 minutes, then measure the surface area. Repeat this
process until the surface area no longer increases. The length of time required to reach this
point is the minimum amount of time required to degas the sample.
12. After the sample is sufficiently degassed, remove the sample tube from the heating station.
Leave the VAC/GAS control knob in the VAC position.
The vacuum fitting for the sample tube may be hot enough to cause burns after the
sample has been heated. Wear the cotton gloves supplied in the accessories kit and
handle the fitting carefully to avoid burns.
Keep the sample tube attached to the vacuum fitting while cooling to prevent
contamination from the ambient atmosphere.
Summary of Contents for 061-00023-00
Page 1: ...VacPrep 061 Operator s Manual 061 42802 01 Feb 2013 ...
Page 2: ... Micromeritics Instrument Corporation 2002 2013 All rights reserved ...
Page 4: ......
Page 46: ...Sample Preparation VacPrep 061 2 26 Feb 2013 ...
Page 56: ......
Page 60: ...Ordering Informatioin VacPrep 061 4 4 Feb 2013 ...