17
Error
Description Potential cause and
remedy
«
Err 6
Triple Mode
There were too many
errors during the
measurement in
Triple mode, making it
impossible to obtain a
final reliable result. Read
through the checklist
for performing reliable
measurements and then
repeat the measurement.
Error
Description Potential cause and
remedy
«
Err 3
»
No pressure
in the cuff
An adequate pressure
cannot be generated
in the cuff. A leak
may have occurred.
Replace the batteries if
necessary. Repeat the
measurement.
«
Err 5
»
Abnormal
result
The measuring signals
are inaccurate and no
result can therefore
be displayed. Read
through the checklist
for performing reliable
measurements and then
repeat the measurement.