41
EN
Guidance and manufacturer’s declaration – electromagnetic immunity –for all EQUIPMENT and SYSTEMS
BT
Guidance and manufacturer´s declaration – electromagnetic immunity
The Model WatchBP Home T(BP3MX1-3T) are intended for use in the electromagnetic environment
specified below. The customer or the user of the Model WatchBP Home T(BP3MX1-3T) should
assure that it is used in such an environment.
Immunity test
IEC 60601 test level Compliance level
Electromagnetic environment -guidance
Electrostatic
± 8 kV contact
± 8 kV contact
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
discharge (ESD)
±2 kV, ±4 kV, ±8 kV, ±2 kV, ±4 kV, ±8 kV,
IEC 61000-4-2
±15 kV air
±15 kV air
Electrostatic
± 2 kV for power
± 2 kV for power
Mains power quality should be that
transient / burst
supply lines 100
supply lines100 kHz of a typical commercial or hospital
IEC 61000-4-4
kHz repetition
repetition frequency environment.
frequency± 1 kV for
input/output lines
Surge
IEC 61000-4-5
± 0.5 kV, ± 1 kV
differential mode line-
line
± 0.5 kV, ± 1 kV
differential mode line-
line
Mains power quality should be that of a
typical commercial or hospital
environment.