Microlife NEB 100 Quick Start Manual Download Page 6

EN

APONORM® COMPACT

6

4.

Malfunction, remedial action

The device does not start.

Ensure that the power plug is properly connected to the power outlet (plugged in).

Ensure that the On/Off switch is in position "1". 

The device may have switched off due to overheating. Wait for the device to cool off and switch it 
on again.

The nebularisation output of the device is insufficient (nebularisation rate too low or none at all).

Make sure that the air tube is properly connected at both ends.

The air tube must not be damaged, compressed, bent, dirty or blocked or clogged. If necessary, 
replace the air tube.

Make sure that the nebuliser kit is properly assembled.

Make sure that the correct quantity of inhalation solution is filled in.

Make sure that all components are well cleaned.

5.

Cleaning and maintenance

Thoroughly clean all accessories after each treatment in order to remove any medication residues and 
potential dirt/contamination. The compressor and the air tube must only be cleaned on the outside 
using a clean and moist cloth. We recommend to disinfect the nebuliser at least once a day.

Always scrub your hands thoroughly before cleaning and disinfecting the accessories.

Never expose the compressor unit to water or heat.

If several persons use one device, each of these persons must have their own complete kit. Change 
it before the treatment of a new person or replace it if it is dirty/contaminated

.

Cleaning:

Clean all nebuliser accessories (except for the air tube) for approx. 5 minutes under warm tap water 
and add a bit of dishwashing liquid (max. 60 °C).

Thoroughly rinse all components. Make sure that all deposits are removed; leave all parts to dry 
completely.

If some parts are very dirty and soiled a clean brush could be used as well.

Disinfection:

Always disinfect the fragmented nebuliser and used components (except the air tube) after regular 
cleaning. Important because only clean materials can be disinfected effectively.

All nebuliser accessories (except for the air tube) can be cleaned using chemical disinfectants. 
Please observe the instructions of the disinfectant manufacturer.

In boiling hot water:

Put the component parts of the nebuliser (except for the air tube, the masks and the compressor unit) 
at least 15 minutes in boiling hot water. Use a clean pot and fill it with enough water. Plastic material 
could melt while touching the hot ground of the pot.

Sterilisation:

All nebuliser accessories (except for the air tube, the masks and the compressor unit) can be 
sterilised using hot steam: up to 121 °C (20 min.) or up to 134 °C (7 min.) EN554/ISO1134.

If available you are also able to use a thermic, usual in trade disinfection tool for baby bottles.

After sterilisation leave all accessories to cool down to ambient temperature. They can be used 
again only after having cooled down. Do not repeat the sterilisation process while the accessories 
are still warm. You can order all replacement parts and accessories from your local pharmacy.

Do not repeat the sterilisation process while the accessories are still warm. You can order all 
replacement parts and accessories from your local pharmacy.

Summary of Contents for NEB 100

Page 1: ...ou have any questions or problems please contact your local pharmacy Your pharmacy will also provide you with any replacement parts required Your pharmacist is the person to contact for all questions...

Page 2: ...attached to both valves 3 Plug the power plug into a power outlet C 4 Remove the top from the nebulizer kit Unscrew the cap of the nebuliser kit by turning it in the direction of the arrow D and lift...

Page 3: ...thetic apparatus or respirator Thoroughly clean the nebuliser kit after each use Make sure that all parts are completely dry and then store the device in a clean and dry place If the device works impr...

Page 4: ...kit by turning it in the direction of the arrow D and lift it off E 5 Add the correct amount of medical liquid to the medication tank F The liquid must not exceed the maximum filling level min 2 ml ma...

Page 5: ...humb and index finger and place it over your mouth and nose Now pull the elastic strap of the mask over your head to fix the mask When donning the mask make sure that the medication tank is kept uprig...

Page 6: ...for the air tube for approx 5 minutes under warm tap water and add a bit of dishwashing liquid max 60 C Thoroughly rinse all components Make sure that all deposits are removed leave all parts to dry...

Page 7: ...elative maximum air humidity Storage temperature 25 to 70 C 13 F to 158 F 10 95 relative maximum air humidity 700 1060 hPa IP Class IP 20 Expected servvice life 5 years Reference to pertinent standard...

Page 8: ...EN APONORM COMPACT 8 3A Health Care S r l Via Marziale Cerutti 90 F Loc San Tomaso 25017 Lonato BS Italy...

Reviews: