9
Microlife BP W1 Basic
EN
9. Safety, care, accuracy test and disposal
m
Safety and protection
Follow instructions for use. This document provides important
product operation and safety information regarding this device.
Please read this document thoroughly before using the device and
keep for future reference.
This device may only be used for the purposes described in these
instructions. The manufacturer cannot be held liable for damage
caused by incorrect application.
This device comprises sensitive components and must be treated
with caution. Observe the storage and operating conditions
described in the «Technical specifications» section.
The cuffs are sensitive and must be handled with care.
Only pump up the cuff once fitted.
Do not use this device if you think it is damaged or notice anything
unusual.
Never open this device.
«ERR
2»
AP
Error signal During the measurement, error signals
were detected by the cuff, caused for
instance by movement or muscle
tension. Repeat the measurement,
keeping your arm still.
«ERR
3»
AQ
No pressure
in the cuff
An adequate pressure cannot be gener-
ated in the cuff. A leak may have
occurred. Check that the cuff is correctly
connected and is not too loose. Replace
the batteries if necessary. Repeat the
measurement.
«ERR
5»
Abnormal
result
The measuring signals are inaccurate
and no result can therefore be
displayed. Read through the checklist
for performing reliable measurements
and then repeat the measurement.*
«HI»
Pulse or cuff
pressure too
high
The pressure in the cuff is too high (over
299 mmHg) OR the pulse is too high
(over 200 beats per minute). Relax for 5
minutes and repeat the measurement.*
Error
Description Potential cause and remedy
«LO»
Pulse too
low
The pulse is too low (less than 40 beats
per minute). Repeat the measurement.*
Error
Description Potential cause and remedy
Summary of Contents for BP W1 Basic
Page 140: ...138 ...