2
8. Safety, Care, Accuracy Test and Disposal
• Safety and protection
• Device care
• Cleaning the cuff
• Accuracy test
• Disposal
9. Guarantee
10. Technical Specifications
Guarantee Card (see Back Cover)
1. Using the Device for the First Time
Inserting the batteries
After you have unpacked your device, first insert the batteries. The
battery compartment
5
is on the bottom of the device. Insert the
batteries (4 x 1.5 V, size AA), thereby observing the indicated
polarity.
Selecting the correct cuff
Microlife offers different cuff sizes. Select the cuff size to match the
circumference of your upper arms (measured by close fitting in the
centre of the upper arm).
Pre-shaped cuffs are optionally available.
Only use Microlife cuffs.
Contact your local Microlife Service if the enclosed cuff
6
does
not fit.
Connect the cuff to the device by inserting the cuff connector
8
into the cuff socket
3
as far as it will go.
2. Checklist for Taking a Reliable Measurement
Avoid activity, eating or smoking immediately before the
measurement.
Sit down on a back-supported chair and relax for 5 minutes.
Keep the feet flat on the floor and do not cross your legs.
Always measure on the same arm
(normally left). It is recom-
mended that doctors perform double arm measurements on a
patients first visit in order to determine which arm to measure in the
future. The arm with the higher blood pressure should be measured.
Remove close-fitting garments from the upper arm. To avoid
constriction, shirt sleeves should not be rolled up - they do not
interfere with the cuff if they are laid flat.
Always ensure that the correct cuff size is used (marking on the cuff).
Fit the cuff closely, but not too tight.
Make sure that the cuff is positioned 1-2 cm above the
elbow.
The
artery mark
on the cuff (ca. 3 cm long bar) must lie over
the artery which runs down the inner side of the arm.
Support your arm so it is relaxed.
Ensure that the cuff is at the same height as your heart.
3. Taking a Blood Pressure Measurement
1. Press the ON/OFF button
1
to start the measurement.
2. The cuff will now pump up automatically. Relax, do not move
and do not tense your arm muscles until the measurement
result is displayed. Breathe normally and do not talk.
3. The cuff fit check
AR
on the display indicates that the cuff is
perfectly placed. If the icon
AR
-A appears, the cuff is fitted
suboptimally, but it is still ok to measure.
4. When the correct pressure is reached, the pumping stops and the
pressure falls gradually. If the required pressure was not reached,
the device will automatically pump some more air into the cuff.
5. During the measurement, the pulse indicator
AQ
flashes in the
display.
6. The result, comprising the systolic
9
and the diastolic
AT
blood
pressure and the pulse rate
AK
is displayed. Note also the
explanations on further display symbols in this booklet.
7. When the device has finished measuring, remove the cuff.
8. Switch off the device. (The monitor does switch off automatically
after approx. 1 min.).
You can stop the measurement at any time by pressing the
ON/OFF button (e.g. if you feel uneasy or an unpleasant
pressure sensation).
This monitor is specially tested for use in pregnancy and
pre-eclampsia. When you detect unusual high readings in
pregnancy, you should measure after a short while again
(eg. 1 hour). If the reading is still too high, consult your
doctor or gynecologist.
Cuff size
for circumference of upper arm
S
17 - 22 cm
M
22 - 32 cm
M - L
22 - 42 cm
L
32 - 42 cm
L - XL
32 - 52 cm