4
Setting the date and time
1. After the new batteries are fitted, the year number flashes in the
display. You can set the year by pressing either the »+«
6
or
the »-«
7
button. To confirm and then set the month, press
enter
1
.
2. Press the »+«
6
or the »-«
7
button to set the month. Press
enter
1
to confirm and then set the day.
3. Follow the instructions above to set the day, hour and minutes.
4. Once you have set the minutes and pressed the time button, the
date and time are set and the time is displayed.
5. If you want to change the date and time, press and hold Date/
Time for approx. 3 seconds until the year number starts
flashing. Now you can enter the new values as described
above.
Selecting the correct cuff
Microlife offers different cuff sizes. Select the cuff size to match the
circumference of your upper arms (measured by close fitting in the
centre of the upper arm).
Only use Microlife cuffs.
Contact your local Microlife Service if the enclosed cuff
AK
does
not fit.
Connect the cuff to the device by inserting the cuff connector
AL
into the cuff socket
3
as far as it will go.
Selecting the user
This device allows to store the results for 2 individual users. In
addition, there is a guest mode in which results are not stored.
Before each measurement
, set the user indicator
BM
for the
intended user: user 1, user 2 or guest mode.
Press and hold the user indicator
BM
for approx. 2 seconds until
the user symbol starts flashing. Now select user 1, user 2 or
guest mode by pressing the user indicator again. Wait 2-
3 seconds to confirm your selection.
The first person to measure should select user 1.
Selecting standard or AFIB/MAM mode
Before each measurement, select standard (single measurement)
or AFIB/MAM mode (automatic triple measurement). In AFIB/MAM
mode, 3 measurements are automatically taken in succession and
the result is then automatically analysed and displayed. Because
the blood pressure constantly fluctuates, a result obtained in this
way is more reliable than when a single measurement is
performed.
When you select the 3 measurements, the AFIB/MAM-symbol
BN
appears in the display.
The bottom, right hand section of the display shows a 1, 2 or 3 to
indicate which of the 3 measurements is currently being taken.
There is a break of 15 seconds between the measurements.
A count down indicates the remaining time.
The individual results are not displayed. Your blood pressure
will only be displayed after all 3 measurements are taken.
Do not remove the cuff between measurements.
If one of the individual measurements was questionable, a
fourth one is automatically taken.
AF detection is only activated in AFIB/MAM mode.
3. Checklist for taking a reliable measurement
Avoid activity, eating or smoking immediately before the
measurement.
Sit down on a back-supported chair and relax for 5 minutes.
Keep the feet flat on the floor and do not cross your legs.
Always measure on the same arm
(normally left). It is recom
-
mended that doctors perform double arm measurements on a
patients first visit in order to determine which arm to measure in
the future. The arm with the higher blood pressure should be
measured.
Remove close-fitting garments from the upper arm. To avoid
constriction, shirt sleeves should not be rolled up - they do not
interfere with the cuff if they are laid flat.
Always ensure that the correct cuff size is used (marking on the
cuff).
Fit the cuff closely, but not too tight.
Make sure that the cuff is positioned 1-2 cm above the elbow.
The
artery mark
on the cuff (ca.3 cm long bar) must lie over
the artery which runs down the inner side of the arm.
Support your arm so it is relaxed.
Ensure that the cuff is at the same height as your heart.
Cuff size
for circumference of upper arm
S
17 - 22 cm
M
22 - 32 cm
M - L
22 - 42 cm
L
32 - 42 cm
L - XL
32 - 52 cm