background image

3

BP A2 Standard

EN

2. Using the Device for the First Time

Inserting the batteries

After you have unpacked your device, first insert the batteries. The 

battery compartment 

5

 is on the bottom of the device. Insert the 

batteries (4 x 1.5 V, size AA), thereby observing the indicated polarity.

Selecting the correct cuff

Microlife offers different cuff sizes. Select the cuff size to match the 

circumference of your upper arms (measured by close fitting in the 

centre of the upper arm).

This device comes with a M or M - L size cuff.

Only use Microlife cuffs.

Contact your local Microlife Service if the enclosed cuff 

6

 does 

not fit.

Connect the cuff to the device by inserting the cuff connector 

7

 

into the cuff socket 

3

 as far as it will go.

3. Taking a Blood Pressure Measurement

Checklist for taking a reliable measurement

1. Avoid activity, eating or smoking immediately before the 

measurement.

2. Sit down on a back-supported chair and relax for 5 minutes. 

Keep the feet flat on the floor and do not cross your legs.

3. Always measure on the same arm (normally left).

4. Remove close-fitting garments from the upper arm. To avoid 

constriction, shirt sleeves should not be rolled up - they do not 

interfere with the cuff if they are laid flat.

5. Always ensure that the correct cuff size is used (marking on the cuff).

Fit the cuff closely, but not too tight.

Make sure that the cuff is positioned 2 cm above the elbow.

The 

artery mark

 on the cuff (ca. 3 cm long bar) must lie over 

the artery which runs down the inner side of the arm.

Support your arm so it is relaxed.

Ensure that the cuff is at the same height as your heart.

6. Press the ON/OFF button 

1

 to start the measurement.

7. The cuff will now pump up automatically. Relax, do not move 

and do not tense your arm muscles until the measurement 

result is displayed. Breathe normally and do not talk.

8. When the correct pressure is reached, the pumping stops and the 

pressure falls gradually. If the required pressure was not reached, 

the device will automatically pump some more air into the cuff.

9. During the measurement, the pulse indicator 

AM

 flashes in the 

display.

10.The result, comprising the systolic 

9

 and the diastolic 

AT

 

blood pressure and the pulse rate 

AK

 is displayed. Note also the 

explanations on further display symbols in this booklet.

11.When the device has finished measuring, remove the cuff.

12.Switch off the device. (The monitor does switch off automatically 

after approx. 1 min.).

4. Appearance of the Heart Arrhythmia Indicator for 

early Detection 

This symbol 

AL

 indicates that certain pulse irregularities were 

detected during the measurement. In this case, the result may 

deviate from your normal blood pressure – repeat the measure-

ment. In most cases, this is no cause for concern. However, if the 

symbol appears on a regular basis (e.g. several times a week with 

measurements taken daily) we advise you to tell your doctor. 

Please show your doctor the following explanation:

5. Data Memory

This device automatically stores the last 30 measurement values.

Viewing the stored values

Press the M-button 

8

 briefly, when the device is switched off. The 

display first shows «

M

» 

AO

 and «

A

», which stands for the average 

of all stored values.

Pressing the M-button again displays the previous value. Pressing 

the M-button repeatedly enables you to move from one stored 

value to another.

Cuff size

for circumference of upper arm

S

17 - 22 cm (6.75 - 8.75 inches)

M

22 - 32 cm (8.75 - 12.5 inches)

L

32 - 42 cm (12.5 - 16.5 inches)

M - L

22 - 42 cm (8.75 - 16.5 inches)

Information for the doctor on frequent appearance of the 

Arrhythmia indicator

This device is an oscillometric blood pressure monitor that also 

analyses pulse frequency during measurement.

The arrhythmia symbol is displayed after the measurement, if 

pulse irregularities occur during measurement. If the symbol 

appears more frequently (e.g. several times per week on 

measurements performed daily) we recommend the patient to 

seek medical advice.

This device does not replace a cardiac examination, but serves 

to detect pulse irregularities at an early stage.

Summary of Contents for BP A2 Standard

Page 1: ...IB BP A2 Standard BY V3 2618 MicrolifeAG Espenstrasse139 9443Widnau Switzerland www microlife com www microlife by www microlife bg www microlife com BP A2 Standard EN 1 RU 8 BG 16...

Page 2: ...Microlife BP A2 Standard BP A2 Standard 3 8 2 1 4 6 7 9 10 11 12 13 16 14 5 15 Guarantee Card www microlife by www microlife bg www microlife com...

Page 3: ...his device uses the same measuring technology as the award winning BP 3BTO A model tested according to the British Hypertension Society BHS protocol Table of Contents 1 Important Facts about Blood Pre...

Page 4: ...per day one in the morning and one in the evening It is quite normal for two measurements taken in quick succes sion to produce significantly different results Deviations between measurements taken by...

Page 5: ...ually If the requiredpressure was not reached the device will automatically pump some more air into the cuff 9 During the measurement the pulse indicator AM flashes in the display 10 The result compri...

Page 6: ...g rechargeable batteries Only use NiMH type reusable batteries Batteries must be removed and recharged when the flat battery symbol appears They should not remain inside the device as they may become...

Page 7: ...The measurement results given by this device is not a diag nosis It is not replacing the need for the consultation of a physi cian especially if not matching the patient s symptoms Do not rely on the...

Page 8: ...of sale made between them 11 Technical Specifications This device complies with the requirements of the Medical Device Directive 93 42 EEC Technical alterations reserved Date of production first 8 dig...

Page 9: ...7 BP A2 Standard EN...

Page 10: ...8 BF 12 Microlife Microlife www microlife by Microlife AG BP 3BTO A BHS 1 2 3 4 5 6 7 Microlife BP A2 Standard RU 1 2 3 4 5 6 7 8 M 9 AT AK AL AM AN AO AP...

Page 11: ...9 BP A2 Standard RU 8 9 10 11 1 o a e 15 4 150 85 120 98 c WHO 2003 100 60 1 100 120 60 80 2 120 130 80 85 3 130 140 85 90 4 140 160 90 100 5 160 180 100 110 6 180 110...

Page 12: ...10 140 90 AP 2 5 4 AA 1 5 Microlife M M L Microlife Microlife 6 7 3 3 1 2 3 4 5 2 3 6 1 7 8 9 AM 10 9 AT AK 11 12 1 4 AL S 17 22 6 75 8 75 M 22 32 8 75 12 5 L 32 42 12 5 16 5 M L 22 42 8 75 16 5...

Page 13: ...11 BP A2 Standard RU 5 30 M 8 M AO A M M 30 31 M CL M CL y 1 M AO M 8 6 AN AN 1 5 2 4 1 5 AA oc a NiMH 7 Microlife 6 600 Microlife 1 4 2 8 ERR 3...

Page 14: ...12 1 9 3 3 ERR 1 ERR 2 ERR 3 ERR 5 HI 299 200 5 LO 40...

Page 15: ...13 BP A2 Standard RU 2 Microlife 10 1 2 3 4 Microlife 5 11 10 40 C 15 95 20 55 C 15 95 375 120 x 87 x 59 20 280 40 200 0 299 1 3...

Page 16: ...14 93 42 EEC 5 4 x 1 5 AA 6 600 920 AA 4 IP20 EN 1060 1 3 4 IEC 60601 1 IEC 60601 1 2 EMC IEC 60601 1 11 12 10000 2...

Page 17: ...15 BP A2 Standard RU...

Page 18: ...16 BF 12 Microlife Microlife www microlife bg Microlife AG BP 3BTO A 1 2 3 4 5 6 Microlife BP A2 Standard BG 1 ON OFF 2 3 4 5 6 7 8 9 AT AK AL AM AN AO AP...

Page 19: ...17 BP A2 Standard BG 7 8 9 10 11 1 15 4 150 85 120 98 mmHg WHO 2003 mmHg 100 60 1 100 120 60 80 2 120 130 80 85 3 130 140 85 90 4 140 160 90 100 5 160 180 100 110 6 180 110...

Page 20: ...18 140 90 AP 2 5 4 x 1 5 V AA Microlife M M L Microlife Microlife 6 7 3 3 1 2 5 3 4 5 2 3 6 1 7 8 9 AM 10 9 AT AK 11 12 1 4 AL S 17 22 6 75 8 75 M 22 32 8 75 12 5 L 32 42 12 5 16 5 M L 22 42 8 75 16 5...

Page 21: ...19 BP A2 Standard BG 5 30 8 M AO A 30 30 31 CL CL 1 M AO 8 6 AN a AN 1 5 2 4 1 5 V AA NiMH...

Page 22: ...20 7 Microlife DC 6V 600 mA Microlife 1 4 2 8 ERR 3 1 9 ERR 1 ERR 2 ERR 3 ERR 5 HI 299 mmHg 200 5 LO 40...

Page 23: ...21 BP A2 Standard BG 3 3 2 Microlife 10 1 2 3 4 Microlife...

Page 24: ...0 40 C 50 104 F 15 95 20 55 C 4 131 F 15 95 375 g 120 x 87 x 59 20 280 mmHg 40 200 0 299 mmHg 1 mmHg 3 mmHg 5 4 x 1 5 V AA e DC 6V 600 mA 920 4 AA IP IP20 EN 1060 1 3 4 IEC 60601 1 IEC 60601 1 2 EMC I...

Page 25: ...23 BP A2 Standard BG...

Reviews: