background image

5

Control program for VT-04

VT-04 control program panel and its elements are shown in the figure below:

COM-port selector

Button to connect/
disconnect the device
and host PC

Activity indicator

Upper point
calibration button

Pane to set speed
under manual mode

Area for hardware
settings indication

Pane to set travel distance
under manual mode

Manual control area

Programmable motion area

Absolute coordinate
of soft 0 position

Current absolute

coordinate of the
carriage

Button to set

current position
as soft 0

Number of remaining cycles

Start/Stop button for
programmable motion

Pane to set number of cycles

Current coordinate

of the carriage
relatively soft 0

Graphic indicator

of soft 0 position

Graphic indicator

of the carriage
current position

Parameters for upward motion

Parameters for
downward motion

[The program start]

Before the control program start, switch the control electronic unit on. After switching the hardware on, wait a couple of

seconds and start the control program. In the folder where the control program was installed, run executable file ‘serial.exe’.

In the control program panel, firstly, select correct COM-port in the port selector (corresponding to ‘USB Serial Port

(COM#)’ in Device Manager). Secondly, press button ‘OPEN RS232’ to connect the device with host PC. About correct
connection system will notify you with label “Device found / dipper control board ready!!!’:

and the port connection button will change color to green with inscription ‘CLOSE RS232’.

Note: If correct connection was not installed, check number of the selected COM-port and repeat the program start procedure

waiting more between the device switching on and the control program start.

Example of the program panel after the device connection is shown below

Reviews: