Page 28
Mounting and Installation
scanCONTROL 25xx
The multifunction socket can be used with either of the following configurations:
Configuration
In1
In2
In3
0
Encoder with index, positive edge works with the index
1
N
A
B
1
Encoder without index, additionally external trigger possible
1
Trigger
A
B
2
External trigger
Trigger
3
External trigger, load up to 4 user modes
Trigger
Mode Bit 0 Mode Bit 1
4
Load up to 8 user modes
Mode Bit 0
Mode Bit 1 Mode Bit 2
5
Transmit in time stamp, (only LLT2510)
Bit 0, ppc 0
Bit 1, ppc 1 Bit 2, ppc 2
6
Frametrigger and encoder
Start container A
B
7
Frametrigger and encoderline A
Start container A
Signal level (switching level):
The signal levels are switchable for all switching inputs together via software between LLL (low-voltage-, TTL
logic) and HLL (high-voltage-, HTL logic):
- LLL level: Low 0 V… 0.8 V, High 2.4 V… 5 V, internal pull-up 10 kOhm to 5 V
- HLL level: Low 0 V… 3 V, High 11 V… 24 V (permitted up to 30 V), internal pull-up 10 kOhm to 24 V
- Pulse duration:
≥
5 µs
i
Use a shielded cable with twisted wires, especially the recommended connection cable PCR3000-x
from the accessories, see Chap.
Connect the cable shield with the potential equalization PE or the connector housing.
1) The encoder input counts each edge. Encoder output typically 4 edges per encoder step.