14
Cleaning
If using the cardboard mouthpieces with the mouthpiece adapter instead of
the SafeBreath filtered mouthpiece (#FM200), it is recommended that the
plastic Mouthpiece Adapter (#PSA1800) is replaced after approximately
250 tests or one month. Re-use can increase the risk of cross
contamination; therefore, it can be cleaned between test using a mild
detergent solution then rinsed with water and left to dry thoroughly.
The device can be cleaned using non-alcohol wipes (see Supplies on page
15
)
. Please be careful not to touch the surface of the sensor or allow any
moisture to touch the sensor because it could cause permanent damage to
the sensor.
Cleaning with products that contain alcohol may result in permanent
contamination of the sensor.
Servicing
If your device requires servicing or repair, please contact Micro Direct, Inc.
directly.
The CO Sensor should be replaced every two years.