HygroCal100 User’s Manual
Michell Instruments
37
APPENDIX A
Appendix A
Technical Specifi cations
Chamber
Generation Range
5 to 95% RH
RH Stability
±0.5%
RH Uniformity
±0.5%
Stabilization Time
Typically <5 min for full stability from step changes of 10% RH
Control Probe
RH Accuracy
±0.8%
Temperature accuracy
±0.2%
Long Term Stability
±1% per year
Electrical Specifi cation
User Interface
4.3" color LCD with touchscreen
Interface with probes
24V excitation voltage, accepts signals: 0-20 mA, 4-20 mA, 0-1
V, 0-5 V, 0-10V
Measurement Units
%RH, temperature in °C, °F
Displayed Resolution
0.1
Data Logging
2Gb internal memory available for log fi les; or 10.6 years storage
at 5 second intervals
Battery (optional)
1500 mAh
Power Supply
24 V DC (100 to 240 V AC, 50/60 Hz adaptor included)
Mechanical Specifi cation
Probe Ports
8 - sensor body diameters 5 to 25mm, accommodated by port
adaptors
Chamber Volume
1050cm
3
approximately
Maximum Probe Insertion Depth 60mm (2.36")
Desiccant Reservoir Capacity
25cm
3
Saturator Reservoir Capacity
25ml
Environmental Conditions
+5 to +40°C (+41 to +104°F)
Dimensions
100 x 250 x 300mm (3.94 x 9.84 x 11.81") h x w x d
Weight
3.2kg (7.05lbs)