I.S. Field Display User’s Manual
18
97321 Issue 2.3, May 2016
OPERATION
3.3.1 Solid State Output
Each alarm has a galvanically isolated single pole solid state switch output which is
shown in
Figure 9. The output is polarized and current will only
fl
ow in one direction.
Ron =
less than 5
Ω
+ 0.7 V
Roff =
greater than 1M
Ω
S
(8, 10)
(8, 10)
(9, 11)
(9, 11)
1M
5Ω
0.7 V
8μH
24nF
+
Figure 9
Equivalent Circuit of Each Alarm Output
3.3.2 Intrinsic
Safety
Each alarm output is a separate galvanically isolated intrinsically safe circuit. The EC-
Type Examination Certi
fi
cate states that for intrinsic safety considerations, under fault
conditions, the output voltage, current, and power at terminals 8 & 9 and 10 & 11 will
not exceed those speci
fi
ed by clause 5.7 of EN 60079-11 for simple apparatus. This
simpli
fi
es system documentation and allows the alarm output terminals 8 & 9 and 10 &
11 to be connected to almost any intrinsically safe circuit protected by a Zener barrier,
or galvanic isolator, providing the output parameters of the circuit do not exceed:
Uo
=
30 V DC
lo = 200
mA
Po = 0.84
W
The maximum equivalent capacitance and inductance between each set of alarm
terminals is:
Ci =
24nF
Li =
8
μ
H
To determine the maximum permissible cable parameters these
fi
gures should be
subtracted from the maximum permitted cable capacitance and inductance speci
fi
ed
by the certi
fi
cate powering the alarm circuit, such as the solenoid driver and switch
transfer galvanic isolators shown in
Figure 10.
Hazardous Area
Safe Area
ppm
V
Easidew PRO I.S. Field Display
P
E
3 4 8 9 10 11 2 1
S
S
S
S
+
+
+
0
9999
4-20 mA
supply
+
0
Solenoid
driver
Alarm 2
Alarm 1
4-20 mA
2-wire
TX
Intrinsically safe
solenoid valve
C tifi d
+
0
Switch
transfer
Pump
Figure 10
Typical Alarm Application
(Shown without recommended screened cables)
Summary of Contents for Easidew PRO
Page 38: ...I S Field Display User s Manual 32 97321 Issue 2 3 May 2016 APPENDIX A...
Page 39: ...I S Field Display User s Manual Michell Instruments 33 APPENDIX A...
Page 40: ...I S Field Display User s Manual 34 97321 Issue 2 3 May 2016 APPENDIX A...
Page 41: ...I S Field Display User s Manual Michell Instruments 35 APPENDIX A...
Page 42: ...I S Field Display User s Manual 36 97321 Issue 2 3 May 2016 APPENDIX A...
Page 43: ...I S Field Display User s Manual Michell Instruments 37 APPENDIX A...
Page 44: ...I S Field Display User s Manual 38 97321 Issue 2 3 May 2016 APPENDIX A...
Page 45: ...I S Field Display User s Manual Michell Instruments 39 APPENDIX A...
Page 46: ...I S Field Display User s Manual 40 97321 Issue 2 3 May 2016 APPENDIX A...
Page 47: ...I S Field Display User s Manual Michell Instruments 41 APPENDIX A...
Page 57: ...I S Field Display User s Manual Michell Instruments 51 NOTES...
Page 58: ...http www michell com...