Easidew Online User’s Manual
Michell Instruments
43
MAINTENANCE
4.3 Fault
Conditions
Message
Displayed
Cause
Action
ErrL
Sensor failure
Check power supply to transmitter. Check
transmitter cable for continuity/damage.
Rectify/replace cable
Instrument failure
Refer to Michell or local representative for
repair
Sbr
Sensor failure or break in
sensor connection
Check transmitter cable for continuity/
damage.
Rectify/replace cable
ErrH
Gas is wetter than +20°Cdp Check gas source supply
Sensor contaminated
Replace/re-calibrate transmitter
outR
Input out of range
Check gas source supply.
Re-calibrate/replace transmitter
rurC
Reverse input connection
Swap input connections from the
transmitter to the monitor