Easidew DryCheck User’s Manual
Michell Instruments
19
OPERATION
3.5
Digital Communication Parameters Set-Up
The default parameters for the Easidew Drycheck are as follows:
Default Address = 1, Baud rate = 9600, Parity = None, Stop bits = 1
To change these parameters, proceed as follows:
Figure 14 shows the operational key sequence.
1. Press
the
P
key once, the display will read
tECH
.
2. Press
the
SET
key and the display will flash between
ConF
and
PinP
.
3. Press
the
key five times, the display will flash between
ConF
and
Corn
.
Set-up instrument address
4. Press
the
SET
key once to display
SAdr
.
5. Press
the
key once to display a flashing 4 digit number. The default
setting is 0001.
6. Use
the
or
keys to give the required new value (e.g. 0002).
NOTE:
The range of possible addresses is between 1 and 247
. Press the
SET
key to accept the new value.
Set baud rate
7.
bAud
will now be displayed. Press the
key once to display a flashing 4
digit number. The default setting is 0003, representing 9600 baud.
8. Use
the
and
keys to give the required new value (the range is 0 to
4). 0 = 1200 baud, 1 = 2400 baud, 2 = 4800 baud, 3 = 9600 baud, 4 =
19200 baud. Press the
SET
key to accept the selected value.
Set parity
9.
Prty
will now be displayed. Press the
key once to display a flashing 4
digit number. The default setting is 0000, representing no parity (none).
10. Use
the
or
keys to give the required new value (the range is 0 to 2).
0 = none, 1 = Odd, 2 = Even. Press the
SET
key to accept the selected
value.
Set number of stop bits
11.
StPb
will now be displayed. Press the
key once to display a flashing 4
digit number. The default setting is 0000, representing 1 stop bit.
12. Use
the
or
keys to give the required new value (the range is 0 - 1)
0 = 1 stop bit, 1= 2 stop bits.
13. Press
the
SET
key to accept the selected value, followed by the
P
key to
return to the Main Display.