(3) warming up the transducer for 3 minutes

(4) using output monitoring meter V1 to measuring the output of signal power

source instrument, set a any input value which is within the measuring range of

the transducer Ur. (for e.g. if input value is 100V, the output is 4-20

mA

, the

expected output value Iz should be calculated as:

Iz = (20mA-4mA)×Ur/100V+4mA

(5) using output monitoring meter V2 to measuring the output voltage U0 of the

transducer, the basic introduced error

γ

of the transducer should be calculated

as:

γ= (Uo-Iz×250Ω)/ [(20mA-4mA)×250Ω]×100

(6)

Repeating (4) (5), if calculated absolute value is less than the given accuracy

value of the transducer, it shows the transducer’s accurate grade is qualified.

Caution:

Caution:

Caution:

Caution:

1.

Pay attention to the auxiliary power information, especially the auxiliary power grade,

and polarity, other wise will damage the product.

2.

pay attention to the wire connection; wrong terminal connection will cause malfunction

of the product and even damage the product;

3.

Don’t dismantle the product, and carry with care to avoiding bump and fall of the

product;

4.

If the product has been using under the environment with strong magnetic field

interference, please pay attention to the shield of input wire, and the output signal wire

should be as short as possible. For product intensive installation, the space between

each product should not be smaller than 10mm.

5.

Only use identified terminals.

6.

There is no lightening strike prevention circuit design in this product. For out door and

hazardous environment using, please add protective alternatives.

7.

This product uses fire prevent ABS crust, its temperature withstand is only limited as

+85

, higher than this limitation will cause the product deformation. Please use and

store carefully.

Summary of Contents for WBV414U01

Page 1: ...01 WBV414U01AC AC AC AC voltage voltage voltage voltage transducer transducer transducer transducer www wb my com wbdz yahoo cn Tel 86 816 2971265 Fax 86 816 2281934 ISO9001 ISO14000 ISO18000 Certified Quality Warranty Any quality problem found in WB series products we offer Three years free charge of repair the products and six months guaranteed free charge of change and return the products ...

Page 2: ...instructions for its installation commissioning and operation However the manual cannot cover all conceivable circumstances or include detailed information on all topics In the event of questions or specific problem do not take any action without proper authorization Contact the appropriate WB technical sales office and request the necessary information Standard Standard Standard Standard applicat...

Page 3: ...w drifting by temperature and wide temperature bearable range etc F F F Function unction unction unction realization realization realization realization schematic schematic schematic schematic drawing drawing drawing drawing Product Product Product Product Dimensional Dimensional Dimensional Dimensional Drawing Drawing Drawing Drawing unit unit unit unit mm mm mm mm E G In 3 1 4 12 11 Out 10 11 _ ...

Page 4: ...mA 3 frequency range 25Hz 5kHz 4 Accuracy 0 5 5 Input Impedance Ri When the input voltage Ux 10V Ri Ux 1kΩ V When the input voltage Ux 10V Ri Ux 10kΩ V 6 Linear Range 0 120 of nominal input 7 Responding Time 300ms 8 Over Load Capacity 2 times of nominal input continuous 9 Output Impedance Output voltage 6V 10 static current 8mA 11 Auxiliary Power 24Vdc 12Vdc 12 Isolation Voltage 2500Vdc for 1 minu...

Page 5: ...tory After correctly wiring it can be powered and used immediately But for further precise signal sampling user need to warm up the product for 3 minutes before use a the auxiliary power requires isolation voltage 2000VAC DC current output ripple 10mV and multiple converters can share with one set auxiliary power together b output signal Iz is designed to follow the impedance load value of 250Ω Wh...

Page 6: ...ed Caution Caution Caution Caution 1 Pay attention to the auxiliary power information especially the auxiliary power grade and polarity other wise will damage the product 2 pay attention to the wire connection wrong terminal connection will cause malfunction of the product and even damage the product 3 Don t dismantle the product and carry with care to avoiding bump and fall of the product 4 If th...

Reviews: