The O2D1 unit emits no sound or start-up oxygen pulse when turned on:

 

1.  Check batteries to make certain they are fresh.

  

  Start-up sound is heard, but no start-up oxygen pulse delivered:

 

1.  Check oxygen cylinder valve is on.

 

2.  Check oxygen supply tube for proper connection.

 

3.  Check oxygen supply tube for obstructions.

 

4.  Check O2D1 outlet tubing for obstructions.

     When using the face mask, no oxygen pulse on inhalation:

 

1.  Make sure the face mask is sealed against the skin.

 

2.  Check for obstructions on the 

O2D1

 outlet tubing.

 

3.  Use only face masks provided by Mountain High Equipment & Supply.  

 

     

NOTE: EDS face masks DO NOT have a plastic dilution bag attached

     Oxygen pulses are delivered, but an alarm is heard at the same time:

 

1.  Check the battery to make certain it is fresh.

 

2.  Check for obstructions in the cannula/mask and tubing.

 

3.  Use only masks and cannulas supplied by Mountain High Equipment & Supply.

     The O2D1 does not trigger at higher altitudes:

 

1.  Try using the Flared-Tip cannula included with the kit (MH part number 00EDS-1084-01)

TROUBLE-SHOOTING 


  

WARNING

 

DO NOT

 

increase or 

decrease length 

of cannula or 

face mask 

supply tubing.

LIMITED WARRANTY

This device is classified as, and is only suitable for use as, a supplementary breathing apparatus (SBA) for aviation use.  It is intended to 
help supply the needed amount of oxygen for persons during flight altitudes where supplemental oxygen is needed.  This device is not 
suitable for any type of life support operations.  This device is not suitable for SCBA (Self Contained Breathing Apparatus), SCUBA (Self 
Contained Underwater Breathing Apparatus) or any medical operations.

Before it is put to use, it is the responsibility of any user who will use this device to become familiar with the operation and safety aspects 
of this device. Using the system improperly could cause failure and lead to possible property damage and/or personal injury. 

Mountain High Equipment & Supply Company assumes no responsibility for property damage, accidents, injury or death that may result 
from  the  misuse  of  this  device/equipment. This  includes  any  use  of  this  device/equipment  outside  the  scope  of  common  sense,  the 
Instruction Manual, inserts and other related documentation.

NOTICE OF NON-LIABILITY

Mountain High Equipment & Supply Company warrants your MH EDS-O2D2 unit against defects in materials and workmanship for two (2) 
years from date of purchase invoice. The warranty is non-transferable. Should any part of the MH EDS-O2D2 become defective within the war-
ranty period return the EDS Unit with a description of what/why it is not functioning and we will repair or replace it, at our discretion, free of 
charge (you pay only shipping to MH).

Return the Unit to:
Mountain High Equipment and Supply Company
RRO Department
625 SE Salmon Avenue Suite 2
Redmond OR 97756-8696

This warranty is non-transferable and only valid if Mountain High Equipment & Supply Company determines that the system and its 
components have not been damaged due to improper use, been submerged in fluids, dismantled or abused. Mountain High Equipment & 
Supply Company reserves the right to determine if repairs are to be done under warranty or at a nominal charge. 

To activate warranty 

coverage, you must complete and return your owner’s enclosed EDS Registration Card.

Summary of Contents for EDS O2D1

Page 1: ...TURES BASIC SAFETY GETTING STARTED MODE CONTROL SWITCH SETTINGS AND MODES OF OPERATION INSERTING OR REMOVING THE BATTERIES 4 2 2 2 3 TROUBLE SHOOTING WARRANTY SPECIFICATIONS REMOVING THE TUBING STORIN...

Page 2: ...it By automatically detecting your pressure altitude the MH EDS O2D1 can be set to start providing oxygen immediately or at a specified altitude and will automatically adjust the oxygen flow as your...

Page 3: ...on the back of the MH EDS O2D1 unit install the 2 AA batteries supplied and replace the battery cover NOTE Batteries fit tightly handle with care See next page for detailed instructions 5 If you are...

Page 4: ...S O2D1 You may mount the MH EDS O2D1 unit to a suitable place using the supplied piece of 3M DUAL LOCK tape Peel the protective backing off one of the rectangles to expose the adhesive and apply it to...

Page 5: ...e F10 setting you will receive the effective flow rate of 5 000 10 000 15 000 ft The F modes are useful for people for whom the standard oxygen supply does not achieve the desired blood oxygen saturat...

Page 6: ...imes with audio beeps once every four 4 seconds until unit detects a valid inhalation event Time before Apnea event is 30 35 seconds This typically occurs for the following reasons 1 The user has quit...

Page 7: ...aratus SCUBA Self Contained Underwater Breathing Apparatus or any medical operations Before it is put to use it is the responsibility of any user who will use this device to become familiar with the o...

Page 8: ...and removed during long term storage Nominal battery voltage 2 875 VDC 40 mv 2 25 ma Idle 100 ma Peak 50 ms max 3 25 ma Average Min start up voltage 2 5 VDC 40 mv Low battery cut out 2 0 VDC 40 mv Red...

Reviews: