METROTEC U15 Series Operating Manual Download Page 9

METROTEC 

 

 

5

 

Device Start-up 

 

5.1 

Switching the Measuring Module on 

The  measuring  module  is  wired  up  in  accordance  with  the  wiring  diagram.  The  module  is 
ready to start when the supply voltage is applied. 
 
 The right LED strip lights up. See Table 1, Figure 1. 
After heating up the sensor, the green LED signals readiness. See Table 1, Figure 2. 
If the current measured value is below the set limit, both LEDs are green. See Table 1, Figure 
3. 
 

 

Table 1: Status display of right LED array 

 

 
Waiting 

for 

sensor 

temperature 
 

 
Sensor  temperature  reached, 
measuring value has exceeded 
the 

limit 

value. 

 

 
Readiness 

and 

measuring 

value have exceeded the limit 
value. 

 

 

 

 

 

 

Figure 1 

Figure 2 

Figure 3 

Ab Bereitschaft liefert das Messmodul am mA-Ausgang den übertragenen Messwert. 
 

5.2 

Measurement 

After the starting routine the device is ready for use and can determine the oxygen content in 
gases. This requires supplying the sample gas according to the sensor type. Alternatively the 
sample gas contains the sensor. See the operating instructions of the respective sensor. 

5.3 

Switching the Measuring Module off 

It is advisable to keep the device in operation continuously. This will avoid the condensation 
of steam in heated sensors which may cause corrosion. 
Should the unit need to be switched off, the energy supply of the measuring module will be 
interrupted. 
See the operating instructions of the respective sensor. 
 

Summary of Contents for U15 Series

Page 1: ...Oxygen Measuring and Control Device U15 Series Measuring module Version 1 6 METROTEC Systeme zur Messung und Regelung von Sauerstoff U15 DIGITAL ...

Page 2: ... Equipment Act Directive 2014 53 EU This device complies with following standards EN 61010 1 EN 61000 6 4 EN 61000 6 2 EN 63000 EN 300220 2 Description of measures taken to assure compliance Quality management system DIN EN ISO 9001 2015 No 12 100 27736 TMS This declaration becomes invalid if changes are made without our consent Kirchheim Teck 27 11 2019 Place Date Signature 2019 METROTEC GmbH Hei...

Page 3: ...n 9 5 2 Measurement 9 5 3 Switching the Measuring Module off 9 6 Measured Value Output 10 6 1 LIN Measured Value Output 10 6 2 LOG Measured Value Output 10 7 Configuration 11 7 1 Enabling Configuration 11 7 2 Correcting the Measured Value 11 7 3 Adjusting the Limit Value 12 7 4 Practical approach 13 7 5 Definition of Analog Outputs 13 7 6 Analog Interfaces 15 7 7 Digital Interfaces 15 8 App for An...

Page 4: ...ensure that the power supplies concur with the details listed on the nameplate All of the covers needed to ensure that the unit cannot be touched when operating must always be fitted You must consider the effects of the overall operation and take the necessary precautions if the unit will be linked up with other equipment and or devices before you switch on Parts and surfaces will occasionally bec...

Page 5: ...the occurance of leakage corrosion condensation etc damage could be done to the equipment and incorrect measurement results be indicated and so it is essential that all parts of equipment be regularly serviced The oxygen sensor and its accessories are subjected to thorough quality control in accordance with DIN ISO 9001 in the course of their manufacture and testing They must only be installated a...

Page 6: ... u15 Digital Designation Sensor Output Alarm Supply U15 Digital Extractive 1 2 230VAC U15 Digital 2 Extractive 2 2 230VAC U15 Digital 24 Extractive 1 2 24VDC U15 Digital 224 Extractive 2 2 24VDC U15 Digital C Inline 1 2 230VAC U15 Digital 2C Inline 2 2 230VAC U15 Digital C24 Inline 1 2 24VDC U15 Digital 2C24 Inline 2 2 24VDC ...

Page 7: ...se oxygen content is to be measured The mechanical construction of the sensor prevents contact between the two gases so that there is no risk of their being intermixed Depending on the application concerned heated or unheated sensors are used Unheated sensors are generally used in furnaces while heated sensors are used for applications where the gas to be measured is at a temperature of less than ...

Page 8: ...n order to implement extensive changes and to display measured values the METROTEC U15 App for Android Smartphones is required 3 3 Sensor Depending on the version the measuring module requires a separate extractive oxygen sensor or an inline sensor to carry out its functions Connecting the sensor See section Connection Diagram as well as separate operating instructions of the sensor 4 General Layo...

Page 9: ...d measuring value have exceeded the limit value Figure 1 Figure 2 Figure 3 Ab Bereitschaft liefert das Messmodul am mA Ausgang den übertragenen Messwert 5 2 Measurement After the starting routine the device is ready for use and can determine the oxygen content in gases This requires supplying the sample gas according to the sensor type Alternatively the sample gas contains the sensor See the opera...

Page 10: ...easured values implicates the assignment of the measured values to either the 0 to 20 mA or the 4 to 20 mA output in logarithmic mode This assignment is permanent and can only be changed on location by means of the METROTEC U15 App for Android Smartphone The output can take on values between 100 and 10 30 The values represent the logarithm of the oxygen partial pressure This presentation helps to ...

Page 11: ...isplay of left LED array No settings possible Adjustment mode for reference point Adjustment mode for Measured value Setting mode for Limit value Figure 1 Figure 2 Figure 3 Figure 4 LED flashes while value is being changed via T3 or T4 7 2 Correcting the Measured Value The current measured value can be corrected if required In this case it is advisable to make the adjustment after a stable measure...

Page 12: ...is flashing When finished the setting is saved by pressing T2 which also serves to leave the setup menu All LEDs in the left array are off See Table 1 Figure 1 7 3 Adjusting the Limit Value The limit value is switched when the current measured value exceeds the current limit value Usually the limit value is within the programmed measuring range An alarm status is signalled at the bottom of the rig...

Page 13: ... a commercially available portable mA measuring device Now the adjustments and settings described above can be carried out After finishing these settings the previously disconnected connections are restored 7 5 Definition of Analog Outputs Switching the analog output from 4 to 20 mA to 0 to 20 mA occurs by means of a jumper between the terminals 12 and 13 Should there be a 2nd analog output it wil...

Page 14: ...A Oxygen 0 5 10 15 20 0 5 10 15 20 mA Oxygen Switching the analog output to linear logarithmic occurs by means of a jumper between the terminals 16 and 17 Should there be a 2nd analog output it will also be switched over in the process Example for logarithmic output 0 to 20 mA at 0 01 100 oxygen ...

Page 15: ...id Smartphone can be connected up As soon as the connection has been established the LED will light permanently Pressing T2 again will switch the Bluetooth connection and the green LED off An exact description is available in the METROTEC U15 App for Android Smartphone operating instructions 9 Specifications Measuring range 100 to 10 35 bar O2 Ambient temperature 0 to 45 degrees Celsius Measuring ...

Page 16: ...mA Output 1 and 2 are electrically isolated from one another Connecting sensors Note Please refer to the operating instructions of the respective sensor Extraktive Sensoren Pin Function 1 Heating 2 Heating 3 Sensor 4 Sensor Inline Sensoren Pin Function 1 Thermocouple 2 Thermocouple 3 Sensor 4 Sensor ...

Page 17: ...easuring output 17 Connection to 16 Connection if logarithmic measuring output U15 Digital 2 Terminal Function Definition 1 230 VAC Supply 2 230 VAC Supply 3 PE Protective Earth Protective Earth 4 mA Measured value output 1 5 mA Measured value output 1 6 mA Measured value output 2 7 mA Measured value output 2 8 Semiconductor relay Readiness 24 VDC 1 A 9 Semiconductor relay Readiness 24 VDC 1 A 10 ...

Page 18: ...measuring output 17 Connection to 16 Connection if logarithmic measuring output U15 Digital 2C Terminal Function Definition 1 230 VAC Supply 2 230 VAC Supply 3 PE Protective Earth Protective Earth 4 mA Measured value output 1 5 mA Measured value output 1 6 mA Measured value output 2 7 mA Measured value output 2 8 Semiconductor relay Readiness 24 VDC 1 A 9 Semiconductor relay Readiness 24 VDC 1 A 1...

Page 19: ...measuring output 17 Connection to 16 Connection if logarithmic measuring output U15 Digital 224 Terminal Function Definition 1 24 VDC Supply 2 24 VDC Supply 3 PE Protective Earth Protective Earth 4 mA Measured value output 1 5 mA Measured value output 1 6 mA Measured value output 2 7 mA Measured value output 2 8 Semiconductor relay Readiness 24 VDC 1 A 9 Semiconductor relay Readiness 24 VDC 1 A 10...

Page 20: ...measuring output 17 Connection to 16 Connection if logarithmic measuring output U15 Digital 2C24 Terminal Function Definition 1 24 VDC Supply 2 24 VDC Supply 3 PE Protective Earth Protective Earth 4 mA Measured value output 1 5 mA Measured value output 1 6 mA Measured value output 2 7 mA Measured value output 2 8 Semiconductor relay Readiness 24 VDC 1 A 9 Semiconductor relay Readiness 24 VDC 1 A 1...

Reviews: