TECHNICAL MANUAL

METRON  ACCUSONIC  PLUS

ULTRASOUND  THERAPY  UNIT

MODEL   AP 100

Prepared by

Metron Medical Australia P/L

Version 1.6   March 2002

Summary of Contents for ACCUSONIC PLUS

Page 1: ...57 Aster Avenue Carrum Downs Victoria Australia 3201 Tel 03 9775 1234 Fax 03 9775 1990 Int 61 3 9775 1234 Int 61 3 9775 1990 TECHNICAL MANUAL METRON ACCUSONIC PLUS ULTRASOUND THERAPY UNIT MODEL AP 100...

Page 2: ...r Metron Accusonic Plus Ultrasound Therapy Unit Model AP 100 is in conformity with the essential requirements and provisions of Council Directive 93 42 EEC is subject to the procedure set out in Annex...

Page 3: ...Should you as the user of this technical manual wish to make any comment about the product or this manual our Authorised Representative within the European Union may be contacted as follows Metron Med...

Page 4: ...7 3 2 Ultrasonic Treatment Applicator s 7 4 Preventative Maintenance Quality Assurance 8 4 1 Calibration and Adjustment 8 4 1 1 Equipment Required 8 4 1 2 Calibration Procedure 8 4 2 Electrical Safet...

Page 5: ...equency 1 1 MHz 10 Output Intensity Continuous Mode 3 0 Watts cm2 20 Maximum Power Intensity Display Accurate to 20 of reading for outputs in excess of 0 2 Watts cm2 Effective Radiating Area 5 0 squar...

Page 6: ...ling between the ultrasonic treatment applicator and the patient On detection of contact loss The LED indicator on the selected applicator handle turns red and after a 2 second delay the treatment tim...

Page 7: ...pment IEC 601 1 1988 Medical electrical equipment Part 1 General requirements IEC 601 2 5 1984 Medical electrical equipment Part 2 Particular requirements for the safety of ultrasonic therapy equipmen...

Page 8: ...il or fax as detailed below Metron Medical Australia Pty Ltd P O Box 2164 CARRUM DOWNS AUSTRALIA 3201 Fax 03 9775 1990 from within Australia 61 3 9775 1990 International The Accusonic Plus generates c...

Page 9: ...er is reduced from that selected to 0 2 Watts cm2 At detection of contact the treatment timer resumes and the ultrasonic output power is restored to the selected value This ensures the patient receive...

Page 10: ...icrocontroller and the tune point of the applicators is stored in the eeprom located in the connector of the treatment applicator L5 monitors supply current level in the output stage Comparing this va...

Page 11: ...ive RF from turning it on The piezoelectric resonator which generates the ultrasound energy is permanently bonded to the cap of the applicator It can be replaced by simply unscrewing the cup and reson...

Page 12: ...display will show a P indicating the next step the alignment of the applicator s output power and contact sense control is ready to be undertaken Place the applicator in the power meter Press the Mode...

Page 13: ...confirm continued operator and patient safety Other mandatory statutory requirements for electrical safety inspections may also apply 4 3 Disassembling Assembling the Unit If required disassembly and...

Page 14: ...5 SCHEMATIC DIAGRAMS 5 1 Main Circuit Board Layout Metron Accusonic Plus __________________________________________________________________ Version 1 6 March 2002 Page 10 of 12...

Page 15: ...5 2 Schematic Diagram Main Printed Circuit Board Metron Accusonic Plus __________________________________________________________________ Version 1 6 March 2002 Page 11 of 12...

Page 16: ...5 3 Schematic Diagram Treatment Head Printed Circuit Board Metron Accusonic Plus __________________________________________________________________ Version 1 6 March 2002 Page 12 of 12...