25
13.
When all the above assembly is finished,
turn on the power switch on the
monochromator. It is located on the lower
surface, on the right-hand side as shown.
The monochromator performs some
initialization tests, which take a moment.
Some noises will be heard as items find
their initial positions. This is normal.
14.
Once the instrument has initialized, the
sample drawer will quietly open as shown.
Do not attempt to push or force the
drawer. It is under instrument control, and
is not designed for manual operation.
Prepare to establish communication from
Vision to the XDS instrument. This is
detailed in Section 6.0, Vision Software.
This completes assembly of the XDS Rapid Liquid Analyzer.