Metrohm Easy PAT D2112 Instruction Manual Download Page 6

Class I equipment is earthed whereas Class II equipment is not earthed (double-insulated),
usually denoted by the symbol 

  

.

Fig . 2

The instrument can be connected directly to a  240V supply using the 13Amp plug fitted on
the supply lead or to a 110V supply, using the adapter supplied.  On power up the EasyPAT
monitors the supply voltage, adjusts its measurement circuits and selects the correct test socket
accordingly.  The display indicates the supply voltage on which tests will be carried out.

Appliances to be tested should be connected into the appropriate test socket and should be
compatible with the supply voltage to the instrument.  If the appliance has an on/off switch, it
should be set to the on  position to ensure that the appliance is fully tested.

1)

EARTH / CONTINUITY (8 AND 25Amps)

For this test a voltage of 6V a.c. is applied between the earth pin of the plug of the
appliance being tested and its exposed metal work, via the wander earth lead.  The
magnitude of the test current is reduced to 8A by a limiting resistor  selected by the
three position  rocker switch.  The high current applied in this test verifies that the
protective earth path will carry the fault current in the event of an insulation breakdown
within the appliance.

2)

EARTH / CONTINUITY (100m A)

The test is similar to that described in (1) except that a constant current of 100m A
d.c. is applied. This reduced current is required to test I.T equipment whose earth
path could be damaged by the higher currents.

5.

CLEANING

Do not use volatile liquids such as thinner or benzene to clean the tester. Wipe the tester with
a soft, dry cloth or moisten  the cloth with a neutral non abrasive detergent and wring the cloth
out thoroughly before wiping over the surface. Ensure that all traces of moisture have evaporated
before  using the tester.

6.

SPARES & ACCESSORIES

1)

PAT Verification Unit

D2101

2)

PAT Extension Lead / IEC Cord Tester

D2102

3)

Mains Adapter - 110V plug to 240V Socket

DFK0056

4)

Flash Gun

DFH0377

5)

16Amp, FF, 11/4”, HBC Ceramic fuss (Front panel)

CEF0107

6)

Instruction Manual

CMR0090

7.

REPAIRS

Please return the tester to:

EDGCUMBE INSTRUMENTS LIMITED
MAIN STREET
BOTHWELL
GLASGOW
G71 8EZ

Be sure to clearly explain the problem in detail.

-3-

-12-

Summary of Contents for Easy PAT D2112

Page 1: ...Main Street Bothwell Glasgow G71 8EZ U K Tel 44 0 1698 852574 Fax 44 0 1698 854442 E mail support metrohm demon co uk CMR0090 Issue 1 2 98 EasyPAT LCD INDICATING PAT MODEL D2112 INSTRUCTION MANUAL Ed...

Page 2: ......

Page 3: ...ENTS i 1 SAFETY RULES 1 2 GENERAL DESCRIPTION 2 5 3 OPERATING INSTRUCTIONS 5 9 4 SPECIFICATIONS 9 4 1 ELECTRICAL 9 11 4 2 MECHANICAL 11 4 3 ENVIRONMENTAL 11 5 CLEANING 12 6 SPARES AND ACCESSORIES 12 7...

Page 4: ...structions before attempting to use this tester Do not exceed the limits of this tester SAFETY CHECK Double check the lead connections before making a test Are you following all the instructions When...

Page 5: ...Y OTHER WARRANTIES EXPRESSED OR IMPLIED The purchaser agrees to assume all liability for any damages and bodily injury which may result from the use or misuse of the product by the purchaser his emplo...

Page 6: ...a limiting resistor selected by the three position rocker switch The high current applied in this test verifies that the protective earth path will carry the fault current in the event of an insulatio...

Page 7: ...at the current flow in the earth lead is monitored to check that no potentially hazardous voltage induced earth leakage paths are created when the appliance is operating 6 FLASH This is an optional te...

Page 8: ...Switch Carefully fit the wander earth lead to any exposed earthed metal on the appliance then press and hold the CONT key The display will show the measured continuity value for the appliance for as l...

Page 9: ...ing by 0 1 and multiply by length for 0 1 e g Maximum cable length for 0 5mm2 on 0 1 setting 2 metres Maximum cable length for 0 5mm2 on 0 2 setting 0 2 0 1 x2 4 metres NOTE During this test high curr...

Page 10: ...lash probe for damage before use Flash voltage is applied i Along the mains lead phase neutral and the earth for a Class I instrument test voltage 1 5kV ii Along the mains lead phase neutral and the f...

Reviews: