![Metrohm 940 Professional IC Vario TWO Manual Download Page 74](http://html1.mh-extra.com/html/metrohm/940-professional-ic-vario-two/940-professional-ic-vario-two_manual_1779626074.webp)
5.9 Servicing the inline filter
■■■■■■■■■■■■■■■■■■■■■■
66
■■■■■■■■
940 Professional IC Vario TWO (2.940.2100)
Removing the filter
Figure 20
Inline filter – Removing the filter
1
Filter housing
Inline filter housing. Part of the accessories
(6.2821.120).
2
Filter screw
Screw for the inline filter. Part of the acces-
sories (6.2821.120).
3
Filter (6.2821.130)
Pack contains 10 pieces.
1 Shutting off the flow
Switch off the high-pressure pump in the software.
2 Removing the inline filter
Unscrew both pressure screws from the inline filter.
3 Unscrewing the filter screw
Use two adjustable wrenches (6.2621.000) to loosen the filter screw
from the filter housing
and unscrew it by hand.
4 Removing the filter