3.15 Connecting the sample degasser (optional)
■■■■■■■■■■■■■■■■■■■■■■
32
■■■■■■■■
940 Professional IC Vario TWO (2.940.2100)
■
PTFE capillary (6.1803.040)
Connecting the sample degasser
CAUTION
If the sample degasser is not used, the inlet and outlet
must
be sealed
with threaded stoppers (6.2744.220).
1 Removing the threaded stoppers
Remove and keep the threaded stoppers (6.2744.220) from the inlet
and outlet of the sample degasser.
2 Connecting the outlet capillary
■
Push a long pressure screw over the loose end of the capillary
connected to Port 1 of the injection valve.
■
Tighten the pressure screw to the outlet of the sample degasser
(labeled
Out
).
Try to ensure the shortest possible connections (shorten the capil-
lary if necessary).
3 Connecting the inlet capillary
■
Push a long pressure screw over one end of the PTFE capillary
(6.1803.040) and tighten the pressure screw to the inlet of the
sample degasser (labeled
In
).