62
■■■■■■■■
884 Professional VA
Activating the calibrator
The calibrator built into the 884 Professional VA guarantees a very high
measuring accuracy for two years after the initial start-up. You can find
more detailed information in the chapter
Calibrator
of the detailed man-
ual.
In
viva
, proceed as follows:
1
Select the 884 Professional VA in the device table of the
Configura-
tion
section of the program.
2
In the device table, click on the
Edit
button and select
Properties...
.
3
Open the
Calibrator
tab.
4
Click on
Activate calibrator
and then on
OK
.
The set-to-work date of the calibrator and the calibrator's serial num-
ber are displayed on the
Calibrator
tab after the dialog window has
been closed and opened again.