4.1 Preparing the ultrafiltration cell
■■■■■■■■■■■■■■■■■■■■■■
10
■■■■■■■■
IC equipment: Inline Ultrafiltration 2 - push mode
4 Installation
4.1
Preparing the ultrafiltration cell
Inserting the filtration membrane
Accessories
■
Ultrafiltration cell 2 (6.02729.110)
■
Filtration membrane 1 (6.2714.020)
■
4 mm hex key (6.2621.030)
■
Plastic tweezers (6.2831.010)
1 Removing the stoppers
■
Remove the 3 green stoppers by hand.
■
Place the ultrafiltration cell on the table with the filtrate chamber
facing upwards. The screws face upwards.
2 Removing the screws
■
Loosen the screws with the hex key.
The screws cannot be removed from the filtrate chamber.