5.2 Carrying out the determination of As(total) according to Application Bulletin 416
■■■■■■■■■■■■■■■■■■■■■■
34
■■■■■■■■
NOTICE
It is recommended to use only reagents of highest quality, e.g. Merck
suprapur
®
or Honeywell/Fluka TraceSelect
®
. The water for preparation
of reagents and for rinsing should have ultrapure quality (resistivity >18
M
Ω·
cm (25 °C), type I grade (ASTM D1193)).
Setup the hardware as described in
Chapter 4.2 Hardware installation
. Make sure that the battery is fully charged.
Establish the connection between software and potentiostat as described
in
Chapter 4.4 Connecting the software with the instrument (page 22)
In daily operation the measuring head module consisting of measuring
head
, scTRACE Gold electrode
, stirrer
and transparent
ring
can stay assembled. Only if the stirrer needs to be replaced, the
transparent ring has to be removed.
Figure 12
Operation of measuring stand