5 Operation and maintenance
942 Extension Module Vario ChS-PP (2.942.0300)
■■■■■■■■
41
Replacing parts of the Metrohm Suppressor Module (MSM)
1 Disconnecting the Metrohm Suppressor Module (MSM) from
the IC system
■
Switch off the instrument.
■
Disconnect all capillaries of the MSM from the IC system.
2 Dismantling the Metrohm Suppressor Module (MSM)
■
Unscrew the union nut
from the housing
.
■
out of the housing together with
the rotor
.
If the rotor gets stuck in the housing, you can push it out as fol-
lows:
Put a pointed object into the slot in the housing and use it to
push out the rotor.
■
Loosen the connecting piece from the rotor with a turning move-
ment.
3 Cleaning the new rotor
■
Clean the sealing surface of the new rotor
using a lint-free cloth.
4 Inserting the new rotor
CAUTION
The rotor may be destroyed during start-up if it is not inserted cor-
rectly.
NOTICE
The adapter (6.2842.020)
(see "Inserting small rotors", page 23)
required in order to insert a small rotor into the suppressor drive.